Treatment
CoolSculpting® Elite
System User Manual
32
CS-UM-CM3-02-EN-D
•
Optional: Gauze soaked in isopropyl alcohol or equivalent wipes
a. Place a towel or other absorbent material into the bucket on top of the control unit.
b. Grasp the applicator to hold it in position; then tap the
OFF
button on the touchscreen to
toggle the vacuum from ON to OFF. The S150 applicator does not use the vacuum
feature and can be removed by detaching the Comfort strap.
Example
: Vacuum OFF
Blue highlight indicates off.
When the vacuum is turned off, the applicator may disengage from the patient.
The applicator could fall and be damaged or cause injury. Grasp the head of the
applicator firmly before turning off the vacuum or releasing the Comfort straps.
c. After you turn off the vacuum, the prompt changes to “Massage” or “New Treatment.”
d. Remove the applicator from the patient and place the applicator head in the bucket on
top of the control unit with the cooling surfaces facing downward.
e. Allow gel to drain into the bucket, or onto a towel or other absorbent material.
f. Massage the treatment area for two minutes: For Dual treatment, the prompt shows 5-
minute delay timer for massage timer for the second side. You can leave the massage
screen when both treatments are completed.
Example
: Massage timer for Applicators A and B for Dual treatment
g. Use towels, gauze or water-based wipes to remove any excess gel from the treatment
area.