System Overview
CoolSculpting® Elite
System User Manual
22
CS-UM-CM3-02-EN-D
Item
Description
Comfort Strap
Minimizes movement of the applicator during treatment.
1. Place bonnet on applicator.
a. Insert applicator in bonnet.
b. Secure bonnet over applicator.
c.
Zip the backside of the bonnet.
2. Position Comfort Strap.
3. Apply the applicator over the desired treatment area.
4. Wrap the Comfort Strap around the applicator.
5.
Secure the Comfort Strap and attach hook tab.
Foam borders
Foam borders minimize movement of the surface applicator during treatment.
The foam borders are for the S150 Applicator only.
Some individuals may be sensitive to
crosslinked ethyl vinyl acetate (EVA) foam or 3M
Double Coated Medical Tape. If a rash develops,
discontinue use and contact a physician.
Note
: Use a new set of borders for each treatment
site. Do not reuse borders.
To apply foam borders for S150 Applicator:
1.
Clean the treatment site with an alcohol wipe.
2. Remove the backing from one pair of foam borders.
3. Apply one pair of foam borders around the treatment site.
4. Repeat above two steps for the other pair of borders.
5. Wipe the treatment site with a pretreatment skin wipe.
6. Repeat above for the other foam borders for Dual treatment
using the S150 Applicator.
7. Apply a gelpad to the treatment site.
Liner
The liner is disposable and provides an interface between the gelpad and applicator.
The liner is for S150 applicator only.
Used liners are considered medical waste. Dispose of used
liners according to your site’s medical waste protocol.
Use a new liner on each treatment
site. Do not reuse liners.
To apply a liner for S150 Applicator:
1. Center the liner above the gelpad.
2. Press the liner onto the gelpad.
3. Working from the center outward,
gently smooth the liner to eliminate
any wrinkles or bubbles.
1