Table 7: Supplies
Item
Description
Card
Provides treatments and profiles for use with the system:
•
Profiles define the number of timed segments of cooling and heating.
•
For the Solo card, each cycle provides a single treatment for one applicator.
•
For the Dual card, each cycle provides a single treatment for both applicators
The card is considered an active device that is inserted in the System when starting treatment.
The card uses software to provide users with the treatment parameters (temperature and time)
with which to conduct treatments.
Filter
The purpose of this filter is to extend the service life of your control unit. Refer to the
Maintenance section for filter replacement instructions.
Gelpad
Provides thermal contact between the applicator and the patient’s skin; mitigates minor
variances in device-to-skin contact.
Note:
Single-use item; use a new gelpad for each
treatment. Reuse of a gelpad may result in
tissue injury. Use a new gelpad each time
you place the applicator on an application
site. Ensure gelpad is appropriate size for
the applicator.
If a gelpad package shows signs of damage, such as leakage, do not use the gelpad.
Store gelpads flat and at room temperature.
Gel
Trap
Fits into the slot of the applicator; prevents
the ingress of gel into the vacuum system.
Single-use item; use a new gel trap for
each treatment.
Refer to Chapter 2 Treatment in the User Manual.
Pretreatment Skin Wipe
Use the Pretreatment Skin Wipe (skin wipe)
to prepare the treatment site before applying
a gelpad by cleaning for 60 seconds.
Single-use item; use a new pretreatment skin wipe for each treatment.