Treatment
CoolSculpting® Elite
System User Manual
28
CS-UM-CM3-02-EN-D
Clean the treatment site with an alcohol wipe.
Clean the treatment site with a pretreatment skin wipe for 60 seconds.
a. Open a gelpad pack and remove the gelpad.
Note:
Gently grasp two corners on a long side of the gelpad
and lift it off the package horizontally.
Example
: grasp two corners on long side of the gelpad
b. Drape the gelpad over the center of the treatment site. Ensure that the gelpad is correct size for
the applicator.
Example
: Apply gelpad on area of treatment for Applicator A
c. Inspect the visible side of the gelpad to ensure that it appears intact.
Use of a damaged gelpad may result in tissue injury. If a gelpad shows signs of
damage, such as tearing, thin spots, or dryness, do not use it.
d. After applying the gelpad, return to the touchscreen and tap
Next
.
The screen showing placement of the applicator over the gelpad appears.
7. Turn on vacuum using single applicator (i.e. Vacuum or Surface Applicator)
Note:
This step focuses on starting a single applicator (i.e. Applicator A).
a. For vacuum applicators, turn on the vacuum by tapping the
ON
button in the upper right corner,
which is next to the vacuum image.
Example
: Vacuum ON
Blue highlight indicates on.
i. When you tap
ON
, the prompt changes to vacuum image with the following text “Waiting for
tissue draw.”