background image

ZScan Manual 

28 

 

3 Positioning 

Targets 

Create a reference by specifying the position of three 

targets. 

 

3 References 

Create a reference with 3 other pre-defined references 

 

In each case, the process is identical: the user has to click on the required features in the 
3D Viewer and define their properties. For example, when using the 3 Planes method, the 

user first selects a plane and must then determine if that plane is the X, Y or Z plane. In 

the case of 3 Positioning Targets, the user must input the position of the target currently 
selected. 

 

In the last step, the user has the option to rotate or to translate the generated reference. 
The result is updated in real time in the 3D Viewer.  

 

 

The units used in the Add Reference Wizard are the units defined 

in ZScan. It defaults to millimeters and can be changed in 

Configure | Options.  

Sensor Calibration 

Even under normal use, the ZScanner 800 may need to have its calibration optimized. This 

tool is used to verify and optimize sensor calibration. To complete the optimization 
process, follow the procedure depicted in the following figure. 

 

 

1.

 

Press the Acquire button to activate the sensor. 

2.

 

Lay the calibration plate on a stable plane surface. Position the scanner at 
approximately 10 cm (4 in) from the calibration plate. 

3.

 

Use the Preview button on the ZScanner 800 to position the laser cross in the white 

cross of the calibration plate. 

4.

 

Press the trigger and slowly move the scanner away from the plate, to an 

approximate distance of 60 cm (24 in) or until the 10 measurements are taken. 

Make sure that the laser cross stays in the white cross of the calibration plate. 

5.

 

Follow recommendations displayed in the Sensor Diagnostic area. 

 

Содержание ZScanner

Страница 1: ......

Страница 2: ......

Страница 3: ...canner 12 Starting ZScan for the first time 13 ZScan 14 General Interface 14 Main Interface Description 15 Configure ZScan 26 Add Reference 27 Sensor Calibration 28 Sensor Configuration 29 Sensor Mana...

Страница 4: ......

Страница 5: ...reflective surface MEDICAL USAGE warn subjects not to stare into the beam Subjects under effect of any medication and young children could possibly have their natural aversion reflex affected and it m...

Страница 6: ...proper authorization is prohibited Distribution of the software without a written authorization from ZCorp is forbidden The Purchaser shall not reproduce or duplicate dissemble decompile reverse engin...

Страница 7: ...you zoom on particular details while scanning Redesigned live surface rendering Enhanced lighting and shading Quick dynamic rendering mode Customize your ZScan experience 5 different 3D viewer control...

Страница 8: ...formation within ZScan New Powerful Installer No need to uninstall ZScan when upgrading Automatic Geomagic detection to install plug in Now possible to keep scanner files when uninstalling Improved Co...

Страница 9: ...ZScanner 800 system includes the following components ZScanner 800 scanner Ergonomic stand Reflective targets Calibration verification plate Carry on case Firewire PCMCIA card Firewire cable Firewire...

Страница 10: ...scanning There are two different ways to acquire these targets either by using the Scan Features mode or by detecting features interactively in the Scan Surface mode The former enables the acquisitio...

Страница 11: ...olution of the volume in order to match the level of detail needed on the object scanned Facets Edition and Multi resolution Facets represent the final output of the scanner which is structured as a c...

Страница 12: ...at can be used to import a positioning model acquired by photogrammetry General Considerations ZScan is a software that takes advantage of modern computer architecture Almost all operations are multi...

Страница 13: ...s all requirements Note that Administrator s privileges are required to complete the installation correctly Insert the ZScan CD ROM in the drive to launch the installation process If the installation...

Страница 14: ...er his name and the organization for which the software is installed It is recommended to install the software for all users so everyone on the machine has access to the software When all these fields...

Страница 15: ...for ZScan and the relevant sensor are selected Also if versions of Geomagic are installed the plugins for this software will be installed It is possible to select not to install one of those component...

Страница 16: ...to use the device with the ZScan application This tool is also accessible through the main ZScan interface and is explained in details later in this manual ZScan is now correctly installed It is poss...

Страница 17: ...card itself The other end of the power supply cable must be plugged into a power outlet and the other end of the FireWire cable must be plugged into the ZScanner 800 The scanner is now correctly conne...

Страница 18: ...located only in the menu Each of these functions will be described and located later The Scan Tree allows the user to select the mode under which the software operates By default the scan mode is sele...

Страница 19: ...process Recording The scanner is currently in the acquisition process Press the trigger to acquire data Disabled The software is currently doing computation and is unable to acquire any data The numbe...

Страница 20: ...possible to select between 5 different mouse configurations The main actions available in each mode are summarized in the table below but at all times the possible mouse actions are listed in the sta...

Страница 21: ...a part The default value is unchecked Lock Viewpoint allows the user to lock the viewer in its current position From that moment it will no longer move automatically when scanning However you can zoo...

Страница 22: ...a slower machine and with a large model If the Edit Facets mode is activated the editing tools will be available in the contextual menu These tools will be described in the Selection Mode section Gen...

Страница 23: ...Start data acquisition Scan Stop Scan Stop data acquisition Scan Reset Scan Reset the current scan The scanner and the scanning parameters will not be reinitialized as in New Session Features Mode Wh...

Страница 24: ...ormation about the total number of features in the current scan and the number of currently observed features The Positioning Features Parameters contains two parameters available to the user the Feat...

Страница 25: ...ften very practical to be able to use another reference or simply add one For example a user may want to scan in a pre defined positioning system to avoid having to align the scan afterward in a third...

Страница 26: ...t possible value that will include all laser curves already acquired regardless of whether they are inside the current volume or not or if the facets have already been deleted Surface Mode Expanding P...

Страница 27: ...2 0 Higher resolution does not necessarily means better results Plane surfaces could end up noisier as more unwanted imperfections could be caught Facets Mode When the Facets mode is selected in the S...

Страница 28: ...functions Facets will be IRREVERSIBLY DELETED When in doubt use the Send facets to Recycle Bin function Facets Mode Expanding Panels There are two expanding panels in Facets mode Facets details and Fa...

Страница 29: ...rts of the 3D model in order to edit parts selectively Selected facets will be displayed in yellow in the 3D Viewer Toolbar Icon Menu Action Edit Selection This button allows the user to choose the di...

Страница 30: ...he Facets Parameters expanding panel Be careful when using the Delete Facets functions Facets will be IRREVERSIBLY DELETED When in doubt leave the facets in the Recycle Bin Recycle Bin Expanding Panel...

Страница 31: ...checked Use My Scans folder ZScan will create a folder in My Documents called My Scans and will use that as the default folder for saving your scans It is possible to set the parameters that ZScan us...

Страница 32: ...its defined in ZScan It defaults to millimeters and can be changed in Configure Options Sensor Calibration Even under normal use the ZScanner 800 may need to have its calibration optimized This tool i...

Страница 33: ...anner to the factory settings via the Sensor Manager Sensor Configuration This tool is used to configure the sensor s laser power and the cameras shutter time according to the type of surface you wish...

Страница 34: ...ting by hitting the Default button It is possible to reset the parameters to what they were before opening the sensor configuration tool by hitting the Reset button Sensor Manager At the end of the in...

Страница 35: ...s manipulation cannot be undone Software Uninstallation ZScan can be uninstalled through the Add and Remove Programs in Windows Control Panel In Windows XP In the Start Menu go to Parameters Control P...

Страница 36: ...keeping his license and configuration files If Yes is selected the folders ZScan Configuration and ZScan Licence will not be removed Help Should you have issues with some of ZScan s functionalities pl...

Страница 37: ...on the object with a minimal distance of 20 mm 0 75 in between them For surfaces with a low curvature the distance between each target can reach 100 mm 4 in Those targets will enable the system to pos...

Страница 38: ...s impossible for ZScan to know in which direction they are The scanner will use the positioning targets pattern to position itself in space and will add the normal after detecting the targets for the...

Страница 39: ...the laser lines detection However if the contrast on the piece is high it is best to scan one part of the object change the sensor configuration and finish the scan with the rest of the scan Please no...

Страница 40: ...get on the plate should be in perfect condition Even only one damaged or misplaced target will produce an Optimization failed message 3 Click on the Default button in order to restore the calibration...

Страница 41: ...nt of a surface already scanned and trigger again or try to redistribute the targets differently on the surface When a temporary malfunction occurs it is indicated by 2 red LEDs located on the upper p...

Страница 42: ...identifies the largest object and removes smaller patches whose size is determined by the value of the slider The bigger patches or the smaller patches that are somehow connected to the main object c...

Страница 43: ...r better precision and accuracy in these zones Decreasing Resolution If some regions in the scan have had their resolution increased it is possible to revert this surface to the original resolution by...

Страница 44: ...become available If the facets were deleted with the Remove Isolated Patches slider it will not be possible to restore them from the Recycle bin Instead you must go back to the Surface Parameters Exp...

Страница 45: ...not detect the targets I am experiencing slow frame rate when scanning My scanner refuses to calibrate I had a Out of memory error and now there is no surface or facets in my session Some holes appear...

Страница 46: ...sold with our systems can be dowloaded by our customers under maintenance through our website Please install these drivers if you are using our laptop under Vista to ensure best possible performances...

Страница 47: ...nstruct the surface at this resolution My surface doesn t appear completely when in Facets view The part you are scanning may be too large or may not be centered in the scanning space Make sure that y...

Страница 48: ......

Страница 49: ...45 Index...

Страница 50: ...36 P PCMCIA 4 Planes 28 Positioning Features 19 35 Positioning Targets 28 Protect Patches 23 R Recycle Bin 5 26 39 Reference Mode 21 Remove Isolated Patches 23 39 Resolution 39 Decreasing 39 Increasin...

Страница 51: ...47...

Отзывы: