Communications
Section
7
YSI Incorporated Drinking Water Monitoring Systems Operations Manual
7-
4
Model: EMS600_
Version:
1.07
Master: 1M!
Sonde: 10617[CR][LF]
The bus master sent a measurement command. The Sonde will take a maximum of 61 seconds to finish the
measurement. Upon completion, it will have 7 sensor data available.
Sonde: 1[CR][LF]
Master: 1D0!
Sonde: 1+17.5+12.05+98.7+8.25+6.45[CR][LF]
Master: 1D1!
Sonde: 1-325+10[CR][LF]
After finishing the measurement, the Sonde sent a service request to indicate completion. The bus master
then sent the "D0" command to retrieve the data. There were 5 data returned. Since 7 readings should be
available, the master continued with "D1" command and received the remaining data. The responses from
"D0" and "D1" commands are:
Temperature:
17.5
Specific
conductance:
12.05
DO %:
98.7
DO mg/L:
8.25
pH (ISE1):
6.45
ORP (ISE2):
-325
Depth:
10