Troubleshooting
Section
6
YSI Incorporated
Drinking Water Monitoring Systems Operations Manual
6-
3
6.3
6.3
6.3
6.3 SENSOR
SENSOR
SENSOR
SENSOR PERFORMANCE PROBLEM
PERFORMANCE PROBLEM
PERFORMANCE PROBLEM
PERFORMANCE PROBLEMSSSS
SYMPTOMS POSSIBLE
CAUSE
ACTION
REF
Probe not properly calibrated
Follow chlorine cal procedures
2.6.1
Membrane not properly installed or
may be punctured
Follow 6572 setup procedure
2.3.1
DO probe electrodes require
cleaning
Follow cleaning procedure. Use
6035 maint. kit
2.10.2
Water in probe connector
Dry connector; reinstall probe
2.3.2
Chlorine Charge too high >30
1. Anode tarnished
Follow cleaning procedure. Use
6035 maintenance kit
2.7
2.10.2
DO probe has been damaged
Replace 6572 probe
8/C
Free Chlorine reading
unstable or inaccurate
Internal failure
Return sonde for service
8
Probe requires cleaning
Follow probe cleaning procedure
2.10.2
Probe requires calibration
Follow cal procedures
2.6.1
3.7.4
pH probe reference junction has
dried out from improper storage.
Soak probe in tap water or buffer
until readings become stable
2.10.2
Water in probe connector
Dry connector; reinstall probe
2.3.2
Probe has been damaged
Replace probe
8/C
Calibration solutions out of spec or
contaminated with other solution
Use new calibration solutions
C
ORP fails Zobell check
Take into account temperature
dependence of Zobell solution
readings
5.3
pH, ORP, chloride,
ammonium, or nitrate
readings are unstable or
inaccurate. Error
messages appear during
calibration.
Internal failure
Return sonde for service
8
Содержание 600DW-B
Страница 3: ......
Страница 239: ...Principles of Operation Section 5 YSI Incorporated Drinking Water Monitoring Systems Operation Manual 5 26...
Страница 251: ...Warranty and Service Information Section 8 YSI Incorporated Drinking Water Monitoring Systems Operations Manual 8 4...
Страница 259: ...Required Notice Appendix B YSI Incorporated Drinking Water Monitoring Systems Operations Manual B 2...
Страница 264: ...EMC Performance Appendix D YSI Incorporated Drinking Water Monitoring Systems Operations Manual D 2...
Страница 268: ...Specifications Appendix E YSI Incorporated Drinking Water Monitoring Systems Operations Manual E 4...
Страница 297: ......