3.7 UPLOADING DATA
3.7 UPLOADING DATA
3.7 UPLOADING DATA
3.7 UPLOADING DATA FROM SONDES
FROM SONDES
FROM SONDES
FROM SONDES
3.7.1 INTRODUCTION
3.7.1 INTRODUCTION
3.7.1 INTRODUCTION
3.7.1 INTRODUCTION
A primary function of the 650 is to serve as an intermediate location for transfer of data stored in deployed 6-
series sondes to a PC without bringing the sondes back to the base facility. In this application, the user
connects the 650 to the sonde at the deployment site via a standard YSI field cable and, using the proper
keypad commands, transfers the data stored in the sonde to the memory of the 650. The sonde is then
redeployed without disturbing the study and the data transferred to a PC on return to the base facility.
3.7.2 UPLOAD PROCE
3.7.2 UPLOAD PROCE
3.7.2 UPLOAD PROCE
3.7.2 UPLOAD PROCEDURE
DURE
DURE
DURE
After attachment of the sonde to the 650 with a YSI field cable, turn on the 650, highlight the
Sonde menu
selection and press
Enter
to display the Main sonde menu. Highlight the File selection and press
Enter
.
NOTE: The upload MUST be made from the File menu of the sonde, not the File menu of the 650.
Select the
Upload
or
Quick upload
(last file) and press
Enter
. Then follow the instructions including
selection of file format – YSI recommends the PC6000 format. Press
Enter
and the display will indicate that
upload of data from sonde to 650 is in progress as shown below. When the upload is complete, the display
will return to the
Sonde File
menu.
After the sonde file has been uploaded to the 650, the data is resident in the flash memory of both sonde and
650. Note that, if you choose to upload files from the sonde in CDF or ASCII format, the files will be
characterized by a .txt extension in the 650 Directory and will NOT be viewable on the 650 display. To view
the files, you must first transfer them from the 650 to your PC and then open them in a spreadsheet.
3.8 USING GPS WITH
3.8 USING GPS WITH
3.8 USING GPS WITH
3.8 USING GPS WITH THE 650
THE 650
THE 650
THE 650
3.8.1 SETTING UP T
3.8.1 SETTING UP T
3.8.1 SETTING UP T
3.8.1 SETTING UP THE GPS
HE GPS
HE GPS
HE GPS----650 INTERFACE
650 INTERFACE
650 INTERFACE
650 INTERFACE
The 650 allows the user to display GPS readings of latitude and longitude from a commercially-available,
user-supplied unit in the Status Bar of the 650 display. The interface of the GPS unit to the 650 is
Содержание 600DW-B
Страница 3: ......
Страница 239: ...Principles of Operation Section 5 YSI Incorporated Drinking Water Monitoring Systems Operation Manual 5 26...
Страница 251: ...Warranty and Service Information Section 8 YSI Incorporated Drinking Water Monitoring Systems Operations Manual 8 4...
Страница 259: ...Required Notice Appendix B YSI Incorporated Drinking Water Monitoring Systems Operations Manual B 2...
Страница 264: ...EMC Performance Appendix D YSI Incorporated Drinking Water Monitoring Systems Operations Manual D 2...
Страница 268: ...Specifications Appendix E YSI Incorporated Drinking Water Monitoring Systems Operations Manual E 4...
Страница 297: ......