Drinking Water Monitoring Systems Operations Manual
2-76
CALIBRATION RECORD – THE GLP FILE
When any sensor is calibrated, most 6-series sondes will automatically create a file in sonde memory that
provides details of the calibration coefficients before and after the calibration. The file will have a .glp
extension and will have the Circuit Board Serial # as the default filename. The file can be viewed by
following the path File|Directory|File Details|View from the Main sonde menu.
Pressing 1-View file will show the calibration record for the sonde. An example is shown below:
------------------File-----------------
1-Directory
4-View file
2-Upload
5-Quick view file
3-Quick Upload 6-Delete all files
Select option (0 for previous menu): 1
Filename Samples
1-BRIDGE1.dat 19
2-BRIDGE2.dat 27
3-UPLAKE.dat 33
4-CLRLAKE2.dat 167
5-DWNLAKE2.dat 31
6 - 00003001.
glp
39
-------------File details--------------
1-View file
2-File:00003001.glp
3-Samples: 39
4-Bytes: 790
Содержание 600DW-B
Страница 3: ......
Страница 239: ...Principles of Operation Section 5 YSI Incorporated Drinking Water Monitoring Systems Operation Manual 5 26...
Страница 251: ...Warranty and Service Information Section 8 YSI Incorporated Drinking Water Monitoring Systems Operations Manual 8 4...
Страница 259: ...Required Notice Appendix B YSI Incorporated Drinking Water Monitoring Systems Operations Manual B 2...
Страница 264: ...EMC Performance Appendix D YSI Incorporated Drinking Water Monitoring Systems Operations Manual D 2...
Страница 268: ...Specifications Appendix E YSI Incorporated Drinking Water Monitoring Systems Operations Manual E 4...
Страница 297: ......