Specifications
Power Accuracy
Item
Specifications
Accuracy
Conditions
(accuracy at 6 months)
Same as the conditions for voltage and current.
Frequency
Accuracy
±
(reading error + measurement range error)
DC
• Voltage input
0.05% of r 0.1% of range
• Current of a 30-A input element
from direct input
• Current of a 2-A input element
0.05% of r 0.1% of range
from direct input
+ 2
µ
A
×
voltage reading
0.1 Hz
≤
f < 30 Hz
0.2% of r 0.3% of range
30 Hz
≤
f < 45 Hz
0.05% of r 0.05% of range
45 Hz
≤
f
≤
66 Hz
• Current sensor input
0.02% of r 0.04% of range
• Current of a 30-A input element from
direct input
• Current of a 2-A input element from
direct input at the 500 mA to 2 A range
• Current of a 2-A input element from
0.05% of r 0.05% of range
direct input at the 5 mA to 200 mA range
66 Hz < f
≤
1kHz
0.05% of r 0.05% of range
1 kHz < f
≤
10 kHz
0.15% of r 0.1% of range
10 kHz < f
≤
50 kHz
0.3% of r 0.2% of range
50 kHz < f
≤
100 kHz
0.014
×
f% of r 0.3% of range
100 kHz < f
≤
500 kHz
0.012
×
f% of r 1% of range
500 kHz < f
≤
1 MHz
(0.048
×
f – 19)% of r 2% of range
* The unit of f in the read error equation is kHz.
•
For temperature changes after zero level compensation or range change, add (voltage influence)
×
(current influence) to the DC accuracy. For details on voltage and current influence, see page 12-5.
•
For self-generated heat caused by current input on an AC input signal, add the following values
to the power accuracy.
•
2-A input element:
0.004
×
I
2
% of reading
•
30-A input element:
0.00002
×
I
2
% of reading
•
For self-generated heat caused by current input on a DC input signal, add the following values to
the power accuracy.
•
2-A input element:
0.004
×
I
2
% of r 6
×
I
2
µ
A
•
30-A input element:
0.00002
×
I
2
% of r 3
×
I
2
µ
A
I is the current reading (A). The influence from self-generated heat continues until the
temperature of the shunt resistor inside the WT3000 lowers even if the current input changes to
a small value.
•
When the data update interval is 50 ms, add 0.1% of reading to all accuracy figures.
•
When the data update interval is 100 ms, add 0.05% of reading to all accuracy figures.
•
All accuracy figures for 0.1 to 10 Hz are reference values.
•
The power accuracy figures for 30 kHz to 100 kHz when the voltage exceeds 750 V are
reference values.
•
The power accuracy figures for 100 kHz to 1 MHz when the voltage (V) exceeds 2.2
×
10
4
/ f
(kHz) are reference values.
•
The power accuracy figures for DC, 10 Hz to 45 Hz, and 400 Hz to 200 kHz when the current
exceeds 20 A are reference values.
•
The power accuracy figures for 200 kHz to 500 kHz when the current exceeds 10 A are
reference values.
•
The power accuracy figures for 500 kHz to 1 MHz when the current exceeds 5 A are reference
values.
12.4 Accuracy
Содержание wt3000
Страница 1: ...Precision Power Analyzer WT3000 IM 760301 01E 8th Edition WT3000 Manual Vol 1 3 ...
Страница 3: ......