4-18
IM 760301-01E
4.6
Selecting the Crest Factor
Procedure
START
STOP
REMOTE
CAL
MOTOR
SET
CURSOR
NULL
MENU
STORE SET
MENU
LOWER ITEM
LOWER FORM
ALL
USER SET
DISPLAY
ITEM & ELEMENT
RESET
SET
PAGE
PAGE
SCALING
HRM SET
MEASURE
SYNC SOURCE
AVG
LINE FILTER
FREQ
FILTER
FILE
MISC
STORE
IMAGE SAVE
HELP
NUMERIC
WAVE
OTHERS
ITEM
FORM
U / I / P
WP/q/
TIME
USER
ELEMENT
UPDATE
RATE
HOLD
SINGLE
INTEG
LOCAL
SHIFT
To exit the menu during operation, press
ESC
located above
the soft keys.
In the procedural explanation below, the phrase “press the
cursor keys
” may be used. This phrase refers to the
procedures for selecting items and entering values and
character strings. For details on the procedures, see section
Cursor keys
1.
Press
MISC
to display the Misc menu.
2.
Press the
Config
soft key to display the Config menu.
• Selecting the Crest Factor
3.
Press
Crest Factor
to select CF3 or CF6.
Explanation
For details, see section 2.3.
• Selecting the Crest Factor
The crest factor is a ratio of the amplitude with respect to the rms value. You can
select CF3 or CF6.
Note
•
If the crest factor is selected, the voltage range and current range are set to the maximum
range on all elements.
•
If the crest factor is set to 6, the measurement conditions of crest factor 5 and higher required
by IEC62018 are met.
•
When measuring waveforms whose crest factor is less than or equal to 3, accurate
measurements can be achieved by setting the crest factor to 3.