• Efficiency Compensation
The power measurement on the secondary side of a power transformer such as an
inverter includes loss caused by the measurement instrument. This loss appears as
error in the efficiency computation. This function compensates for this loss. For
details, see appendix 8.
• Compensation for the Two-Wattmeter Method
In the two-power wattmeter method, an error results when current flows through the
neutral line. This function computes the current the flows through the neutral line for
measurements using the two-wattmeter method with a three-phase, three wire (3V3A)
system and adds the compensation value to the measured power. For details, see
appendix 8. This function can be set on models with the delta computation option.
Note
•
If the current input is an external sensor input (EXT) and the wiring compensation is U-I
wiring, wiring compensation is not performed. This is because the instrumental loss of the
current measurement is unknown.
•
If the current input is an external sensor input (EXT), efficiency compensation is not
performed. This is because the instrumental loss of the current measurement is unknown.
•
Efficiency compensation and two-wattmeter method compensation are not performed in
wide bandwidth harmonic measurement and IEC harmonic measurement modes on
models with the advanced computation (/G6) option.
•
Wiring compensation, efficiency compensation, and two-wattmeter method compensation
are not performed in voltage fluctuation and flicker measurement, and cycle-by-cycle
measurement modes. Settings for these compensation functions are invalid.
5.8 Setting the Wiring, Efficiency, and Two-Wattmeter Method Compensations