8-6
IM 760301-01E
Explanation
• Selecting the Items to Be Stored
Select the item to be stored from the following:
• Numeric
Numerical data are stored.
• Wave
Waveform display data are stored.
• Wave
Numeric data and waveform display data are stored.
• Selecting the Numeric Data to Be Stored
Select the items for storing the numeric data.
• Installed elements and wiring units are applicable for the selection.
• The items to be stored can be selected from the items that are indicated in
“Measurement Function Types during Normal Measurement” (see section 2.2),
“Measurement Function Types for Motor Evaluation Function (Motor Version)” (see
the
Expansion Function User’s Manual IM760301-51E
), and “Measurement
Function Types for Harmonic Measurement (Option)” (see the
Expansion Function
User’s Manual IM760301-51E
).
• Selecting the Waveform Display Data to Be Stored
• Select the items for storing the waveform display data.
• Only the input signals of the installed elements are applicable for the selection.
• For models with the motor evaluation function (motor version), the input signals of
Speed and Torque are also selectable items.
Note
•
There are limitations on the setting and execution of the storage operation in measurement
modes other than normal measurement mode. For details, see section 8.1.
•
For the meanings of the measurement function symbols that are displayed, see section 2.2,
“Measurement Functions and Measurement Periods,” 2.5, “Computation,” 2.6, “Integration,”
appendix 1, “Symbols and Determination of Measurement Functions,” and
Expansion
Function User’s Manual IM760301-51E
•
For details on the wiring units expressed as
Σ
A and
Σ
B, see section 4.1, “Selecting the Wiring
System.”
•
[-------] (no data) is stored in places where there is no numeric data.
•
If there are no integrated values because integration is not performed, [-------] (no data) is
stored. [-------] (no data) is also stored in places where there is no integration time.
•
The maximum order for which the harmonic data is saved is the maximum harmonic order to
be measured that is specified in the harmonic measurement (option) menu. [-------] (no data)
is stored for harmonic orders without data.
•
To store waveform display data, set the display mode to Wave, Wave, Wave+Bar,
or Wave+Trend and display the waveform on the screen. If the display mode is not set to one
of these modes, “NAN” is stored for the waveform data.