10.1 Zero-Level Compensation
Procedure
START
STOP
REMOTE
CAL
SENSOR RATIO
MEASURING
RMS
MEAN
DC
RMEAN
CURRENT RANGE
VOLTAGE RANGE
MEASURING
RMS
MEAN
DC
RMEAN
ALL
INPUT INFO.
LOWER ITEM
LOWER FORM
ALL
USER SET
DISPLAY
ITEM & ELEMENT
2
3
4
COMPEN
ELEMENT
ELEMENT
1
WIRING
MODE
AUTO
MODE
AUTO
EXT SENSOR
NUMERIC
WAVE
OTHERS
ITEM
FORM
U / I / P
WP/q/
TIME
USER
ELEMENT
UPDATE
RATE
HOLD
SINGLE
INTEG
LOCAL
SHIFT
RESET
SET
PAGE
PAGE
Cursor keys
To exit the menu during
operation, press
ESC
located above the soft
keys.
In the procedural
explanation below, the
phrase “press the
cursor
keys
” may be used. This
phrase refers to the
procedures for selecting
items and entering values
and character strings. For
details on the procedures,
see section 3.14.
Press
SHIFT
+
SINGLE (CAL)
. Zero-level compensation is executed.
Explanation
Zero-Level Compensation
This function is used to create a zero input condition using the internal circuit of the
instrument and setting the level at that point to zero level. Zero-level compensation
must be performed to meet the specifications of this instrument (see chapter 12).
• Pressing CAL executes the zero-level compensation.
• Zero-level compensation is automatically performed after changing the
measurement range or input filter.
Note
•
For making accurate measurements, it is recommended that zero-level compensation be
performed after warming up the instrument for at least 30 minutes. In addition, the ambient
temperature needs to be within the specifications (see chapter 12) and stable.
•
If the measurement range and input filter are not changed over a long period of time, the zero
level may change due to the changes in the environment surrounding the instrument. It is
recommended that zero-level compensation be performed in this case.
•
There are functions that automatically perform zero-level compensation during integration.
See section 5.13.
Chapter 10 Other Functions