<6. Functions>
6-12
IM 12K01B02-01E
6.5.2
Setting of Decimal Place for each Component
MEAS.
screen
MAINTENANCE
screen
[MAINT.]
[MEAS.SETTING]
Refer to “6.5 Setting Screen Display for each Measuring Component”
MEAS. SETTING for each
measuring component screen
MEAS. SETTING
screen
[UV][VIS][COD][TURB]
The decimal place for each measuring component is set here.
1. Press the button for DECIMAL PLACE on the MEAS. SETTING for each measuring
component screen.
The DECIMAL PLACE screen appears.
The screen shows an example for UV.
DECIMAL
PLACE
2. Enter a numeric value for the decimal place and press [SET].
The screen returns back to the MEAS. SETTING for each measuring component screen.
The decimal place setting for each measuring component is complete.
Perform the same procedure to set the measuring range for the other components.
CAUTION
The decimal place for UV-αVIS is the same as that for UV.
6.5.3
Unit Setting for UV, VIS, and UV-αVIS
MEAS.
screen
MAINTENANCE
screen
[MAINT.]
[MEAS.SETTING]
Refer to “6.5 Setting Screen Display for each Measuring Component”
UV MEAS. SETTING
screen
MEAS. SETTING
screen
[UV]
The units for UV, VIS, and UV-αVIS are set here. Select “Abs” when indicating UV, VIS, and
UV-αVIS in a cell length conversion of 10 mm, and select “m
-1
” when indicating UV, VIS, and
UV-αVIS in a cell length conversion of 1 m (the indication of “m
-1
” is the hundredfold of the
indication of “Abs” ).
When the unit is changed, the range that can be measured simultaneously is also changed.