<9. Troubleshooting>
9-4
IM 12K01B02-01E
Item
Alarm
Causes
Measures
Refer to:
UV
ZERO CAL.
Incompatible calibration solution. Make sure to use distilled water for
the calibration solution.
Contamination of measuring cell.
Clean the measuring cell.
Replace the wiper blade rubber
when it is deteriorated.
"8.2.1 Cleaning Method for
Measuring Cell"
"8.2.2 Replacement of Wiper
Blade Rubber"
Failure of parts in the detector.
Replace the faulty parts.
Service call
SPAN CAL.
Incompatible calibration solution.
Make sure to use the exclusive
calibration solution, and distilled
water as the dilution water.
"5.4.3 Common Span
Calibration"
Contamination of measuring cell.
Clean the measuring cell.
Replace the wiper blade rubber
when it is deteriorated.
"8.2.1 Cleaning Method for
Measuring Cell"
"8.2.2 Replacement of Wiper
Blade Rubber"
Failure of detector parts.
Replace the faulty parts.
Service call
VIS
ZERO CAL.
Incompatible calibration solution. Make sure to use distilled water for
the calibration solution.
Contamination of measuring cell.
Clean the measuring cell.
Replace the wiper blade rubber
when it is deteriorated.
"8.2.1 Cleaning Method for
Measuring Cell"
"8.2.2 Replacement of Wiper
Blade Rubber"
Failure of detector parts.
Replace the faulty parts.
Service call
SPAN CAL.
Incompatible calibration solution.
Make sure to use the exclusive
calibration solution, and distilled
water as the dilution water.
"5.4.3 Common Span
Calibration"
Contamination of measuring cell.
Clean the measuring cell.
Replace the wiper blade rubber
when it is deteriorated.
"8.2.1 Cleaning Method for
Measuring Cell"
"8.2.2 Replacement of Wiper
Blade Rubber"
Failure of detector parts.
Replace the faulty parts.
Service call