<6. Functions>
6-3
IM 12K01B02-01E
6.4
Maintenance - Setting
All maintenance settings are performed here.
6.4.1
Signal Allocation
MEAS.
screen
MAINTENANCE
screen
[MAINT.]
[SETTING]
Refer to "6.3 Maintenance Screen Display"
SIGNAL ALLOCATION
screen (1/3)
SETTING 1/2
screen
[SIGNAL INPUT SETTING]
The output terminals for analog output and contact output are allocated here.
1. Press [SIGNAL ALLOCATION] on SETTING 1/2 screen.
The SIGNAL ALLOCATION screen appears.
[SIGNAL
ALLOCATION]
[ ]
[ ]
2. Output signal is allocated.
Setting item
Description
Initial setting
ANALOG OUTPUT A-1 Sets to output either one of the following measurement component to
the analog output terminal from A-1 to A-3 Setting range:
UV, VIS, UV-VIS, COD, TURB, NONE
COD
ANALOG OUTPUT A-2
TURB
ANALOG OUTPUT A-3
UV
SIGNAL OUTPUT O-1
Sets to output either one of the following items to the output terminal of
contact output from O-1 to O-4 Setting range:
TOTAL ALARM, MAX COD, MAX TURB, MIN COD, MIN TURB, LAMP
ERR., NO SAMPLE, MOTOR ERR., ANALYZE ERR., WARM-UP
TOTAL ALARM
SIGNAL OUTPUT O-2
MAX COD
SIGNAL OUTPUT O-3
MAX TURB
SIGNAL OUTPUT O-4
NO SAMPLE
SIGNAL OUTPUT O-5
Outputs the following item to the output terminal O-5 of contact output.
POWER OFF (cannot be changed)
POWER OFF
(fixed)
SIGNAL OUTPUT O-6
Outputs the following item to the output terminal O-6 of contact output.
MAINTENANCE (cannot be changed)
MAINTENANCE
(fixed)
Refer to “7 External Input and Output”