<5. Calibration>
5-13
IM 12K01B02-01E
4. Pour the residual individual span solution into the cleaning bottle.
Cleaning bottle
Residual individual (UV) span solution
When performing span calibration, the calibration work can be performed smoothly if the
preparation through step 4 is performed together with the preparation for the common zero
calibration.
5. Take the detector from the zero calibration tank, and clean the cell unit using the cleaning
bottle containing the individual span solution.
Clean the cell unit using the cleaning bottle containing
the individual (UV) span solution.
CAUTION
Do not spill any solution on the base of the cell rotary shaft (V-ring seal part).
6. After cleaning the cell unit with the individual span solution, dip in the calibration tank.
Allow the solution to adhere by shaking from side to side
for several times.
7. Press the button for [SPAN VALUE] for UV on the UV/VIS CAL. screen.
The screen changes to the UV SPAN screen.
[X.XX]
NOTE
[CLEAR]: Clears all input numeric values.
[BACK]: Clears the rightmost number.