<6. Functions>
6-15
IM 12K01B02-01E
3. Enter the value of “a” with the numeric keypad, and press [SET].
Input range: -9999.9 to 9999.9
(When the decimal place is set to 1 digit, the decimal place increases according to the
setting digits.)
The screen returns back to CORRECTION FACTOR screen for each measuring
component.
4. Press [b=x.xxx].
The CORRECTION FACTOR (b) screen for each measuring component appears.
NOTE
[CLEAR]: Clears all input numeric values.
[BACK]: Clears the rightmost number.
5. Enter the value “b” with the numeric keypad, and press [SET].
Input range: 0.1 to 9999.9
(When the decimal place is set to 1 digit, the decimal place increases according to the
setting digits.)
The screen returns back to the y = a + bx screen for each measuring component.
The setting of the conversion factor and measurement correction factor are complete.
6.5.6
Setting of MAX Alarm Value for COD and TURB
(turbidity)
MAX ALARM
VALUE
1. Here describes how to set the upper limits of the COD and TURB alarm values.
Press the button of the measuring component on the COD or TURB MEAS. SETTING
screen.
The MEAS. SETTING screen for each measuring component appears.