<6. Functions>
6-16
IM 12K01B02-01E
2. Press the button for MAX ALARM VALUE.
The MAXIMUM ALARM VALUE screen for each measuring component appears.
The screen shows an example of MAX ALARM VALUE for COD.
MAX ALARM
VALUE
3. Enter the upper limit value with the numeric keypad, and press [SET].
Input range: -9999.9 to 9999.9
(When the decimal place is set to 1 digit, the decimal place increases according to the
setting digits.)
NOTE
[CLEAR]: Clears all input numeric values.
[BACK]: Clears the rightmost number.
The screen returns back to the MEAS. SETTING screen for each measuring component.
6.5.7
Setting of MIN Alarm Value for COD and TURB (turbidity)
MEAS.
screen
MAINTENANCE
screen
[MAINT.]
[MEAS.SETTING]
Refer to “6.5 Setting Screen Display for each Measuring Component”
COD MEAS. SETTING screen
TURB MEAS. SETTING screen
MEAS. SETTING
screen
[COD][TURB]
Here describes how to set the lower limits of the COD and TURB alarm values.
1. Press the button of the measuring component on the COD or TURB MEAS. SETTING
screen.
The MEAS. SETTING screen for each measuring component appears.
2. Press the button for MIN ALARM VALUE.
The MINIMUM ALARM VALUE screen for each measuring component appears.
The screen shows an example of MINIMUM ALARM VALUE for COD.
MIN ALARM
VALUE
3. Enter the lower limit value with the numeric keypad, and press [SET].
Input range: -9999.9 to 9999.9
(When the decimal place is set to 1 digit, the decimal place increases according to the
setting digits.)
The screen returns back to the MEAS. SETTING screen for each measuring component.