<5. Calibration>
5-3
IM 12K01B02-01E
2. Press the measurement items to be calibrated on the CALIBRATION screen.
The CALIBRATION screen for each measurement item is displayed.
Calibration item
[UV/VIS]
Example when the calibration item is UV/VIS:
3. Perform calibration by following the calibration procedure for each measurement item.
5.4
Calibration of UV/VIS
There are three kinds of calibration method for UV/VIS: common zero calibration, common span
calibration, and individual span calibration.
Refer to “5.1 Calibration Pattern and its Cycle”
5.4.1
How to Prepare Calibration Solution
Use distilled water for the zero calibration solution, and exclusive calibration solution for the
common span calibration solution.
The calibration solution ampoule (H calibration solution (blue label)) for 1 Abs
(100 m
-1
) is attached to this unit. An attached calibration solution ampoule covers the
measurement range up to 1 Abs. Full scale can be covered from 0.1 Abs to 5.0 Abs (10 m
-1
to
500 m
-1
) by using the 2 or 4 calibration solution ampoules.
Select the measurement full scale you use from the following, and compound the calibration
solution.
Measurement full
scale range
Qty. of calibration
solution ampoule
Concentration of calibration
solution when diluted with 2 L
Calibration solution
0.1 Abs to 1.0 Abs
(10 m
-1
to 100 m
-1
)
1
UV: 0.690 Abs (69.0 m
-1
)
VIS: 0.880 Abs (85.0 m
-1
)
1.1 Abs to 2.5 Abs
(110 m
-1
to 250 m
-1
)
2
UV: 1.380 Abs (138.0 m
-1
)
VIS: 1.760 Abs (170.0 m
-1
)
2.6 Abs to 5.0 Abs
(260 m
-1
to 500 m
-1
)
4
UV: 2.760 Abs (276.0 m
-1
)
VIS: 3.520 Abs (340.0 m
-1
)
H calibration
solution (blue label)
0 to 1 Abs for
calibration
Also, our calibration solution ampoule (H II calibration solution (green label)) for 2 Abs (200 m
-1
)
is also available. If using in the measurement full scale of 1.1 Abs (110 m
-1
) or over, it is more
economical to buy H II calibration solution ampoules as the number of ampoules used is fewer.
Measurement full
scale range
Qty. of calibration
solution ampoule
Concentration of calibration
solution when diluted with 2 L
Calibration solution
1.1 Abs to 2.5 Abs
(110 m
-1
to 250 m
-1
)
1
UV: 1.380 Abs (138.0 m
-1
)
VIS: 1.760 Abs (170.0 m
-1
)
2.6 Abs to 5.0 Abs
(260 m
-1
to 500 m
-1
)
2
UV: 2.760 Abs (276.0 m
-1
)
VIS: 3.520 Abs (340.0 m
-1
)
H II calibration solution
(green label)
0 to 2 Abs
for calibration
l
Common zero calibration solution
Preparation:
Distilled water: 2 L
Make sure to use the distilled water.