SingleN:
The instrument acquires signals each time the trigger conditions are met until a
specified number of signals have been acquired, stops acquisition, and displays
all of the acquired signals. If no triggers occur, the display is not updated.
On Start:
Regardless of the trigger settings, when you press the START key, the instrument
updates the displayed waveforms once and stops signal acquisition. If the time
axis is set to a value that would cause the display to switch to roll mode, roll
mode display will be enabled. When data has been acquired up to the amount
specified by the set record length, the waveform display stops.
4.1 Setting the Trigger Mode