Setting the Display Range in Recorder Mode (Upper/Lower)
3.
After step 2 on page 2-3, tap
Upper/Lower
, or press the soft key.
Set the upper limit.
Set the lower limit.
Note
The upper and lower limits cannot exceed 10 times the measurement range. If changing the
measurement range causes the upper and lower limits to exceed 10 times the measurement range,
the upper and lower limits are automatically adjusted.
Setting the Adjustment (DC offset cancellation, gain adjustment)
3.
After step 2 on page 2-3, tap
Adjustment
, or press the soft key.
Acquires the DC offset value
Turns DC offset canceling on or off
Set the gain adjustment (0.9500 to 1.0500).
Skew adjustment (-1.0 µs to 1.0 µs).
(On modules whose maximum sample rate is
10 MS/s or more)
Note
• When DC offset canceling is set to ON, you cannot adjust the gain.
• When gain adjustment is enabled, DC offset canceling cannot be set to ON. If you acquire the DC offset
value when gain adjustment is enabled, gain adjustment will be disabled, and DC offset canceling will be
set to ON.
Setting the Sample Rate (SampleRate, Scope mode)
In Scope mode, you can set the sample rate for each channel.
You can set the sample rate as high as the sample rate determined by the display record length
and T/div settings in the waveform acquisition conditions.
If the sample rate determined by the display record length and T/div settings exceeds the maximum
sample rate of a module, you can set the sample rate as high as the maximum sample rate of that
module.
For example, if modules with sample rates at 10 MS/s or higher are installed and the sample rate
determined by the display record length and T/div settings exceeds the maximum sample rate of a
module, you can set the sample rate as high 10 MS/s.
This is useful when you do not want to acquire unneeded data.
Setting the Sample Interval (Sample Interval, Recorder mode)
In Recorder mode, you can set the sample interval for each channel.
You can set the sample interval as high as the sample interval set on the Acquire menu.
This is useful when you do not want to acquire unneeded data.