![YOKOGAWA ScopeCorder DL950 Скачать руководство пользователя страница 103](http://html.mh-extra.com/html/yokogawa/scopecorder-dl950/scopecorder-dl950_user-manual_911350103.webp)
4-15
IM DL950-02EN
4.10 Triggering on an Edge On A Trigger (Enhanced)
This section explains the following settings for triggering on an Edge On A trigger:
Applicable in Scope Mode
• Trigger source
• State condition
• State condition achievement condition
• Edge detection condition
• Trigger conditions (trigger level, hysteresis)
► “Edge On A Trigger (Enhanced)” in the Features Guide
Trigger Menu
1.
On the waveform screen’s
MENU
, tap
TRIGGER
>
MENU
, or press
TRIG.MENU
. A trigger
menu appears.
Select Enhanced.
Set Type to Edge On A.
See “Setting the State and Trigger
Conditions.”
Set the trigger position, trigger delay, and hold-off
► sec. 4.2
Set the trigger mode. ► sec. 4.1