22-7
IM DL950-02EN
22.3 Starting and Stopping Waveform Acquisition
Items Common to Scope Mode and Recorder Mode
• Starting and stopping waveform acquisition
► “Synchronizing Multiple Instruments (/C50 option)” in the Features Guide
Waveform Acquisition (START and STOP)
Perform the operation on the main unit.
1.
On the main unit waveform screen’s
MENU
, tap
START/STOP
, or press
START/STOP
.
Waveform acquisition starts or stops on the main unit and sub units.
Waveform acquisition is in progress when the START/STOP key is illuminated.
Setting Items That Give Priority to the Settings of the Main Unit
When multiple DL950s are connected, the following settings on the main unit take priority. The
settings on the sub units are overwritten with the settings on the main unit.
• Record length
• Time base, T/Div
• Acquisition mode
• Trigger mode
• Number of measurements
• SSD recording settings
• Dual capture settings
• Trigger position, trigger delay
• Hold-off time
• Action trigger settings
• Accumulation display
• Current history number
• Search feature
• Go-Nogo settings
• Save operation
The following features will not work when the units are linked.
• Averaging function
• Single-N trigger, line trigger, wave window trigger, sub units’ external trigger
• Enhanced trigger other than OR
• History search
• History averaging
• All history record display
• Waveform acquisition start using power-on action
• Remotely stopping waveform acquisition