9-6
IM DL950-02EN
Searching by Automatically Measured Parameter (Search Setup)
2.
Tap
Search Mode
, or press the soft key. On the menu that appears, Search Mode to
Parameter
.
3.
Tap
Search Setup
, or press the soft key. The following screen appears.
Register the search parameter (Pram1-Param4).
Set the search condition (OFF, IN, OUT).
See “Sitting the Source Waveform and Parameters.”
Set the upper and lower search condition limits.
Set the search logic (AND, OR).
(For Param1 to Pram4)
Measurement time period of the parameter (left edge –
Time Range1; right edge – Time Range2).
(For Param1 to Pram4)
For all items other than Logic and Time Range1, 2, set for each Param.
Sitting the Source Waveform and Parameters (Source)
4.
Press the
Source
soft key. The following menu appears.
Select the source waveform (CH1, CH16).
Set the waveform parameter type. ► sec 11.1
Executing a History Waveform Search
5.
After you set all the search conditions, close the menu to return to the HISTORY menu.
6.
Tap
Execute
, or press the soft key. The instrument searches for the search conditions.
The soft key changes to Abort.
To abort a search, tap
Abort
, or press the soft key.
9.2 Searching History Waveforms