11.4 Appendix 4: Factory defaults & setting limits SC
The following table shows the details of the factory defaults of the SA11 and their setting
limits. Setting can be done by HOST system but some setting limits are restricted by
limitation of HOST.
Table A4.1: Factory defaults & setting limits SC
Item
Factory
Setting limits
Unit
default
Lower limit
Upper limit
Measurement settings
manual_temp
+25 / +77
-40 / -40
+260 / +500
°C / °F
temp_ref
+25 / +77
-40 / -40
+260 / +500
°C / °F
temp_offset
0 / +32
-10 / +14
+10 / +50
°C / °F
external_temp
NaN
-40 / -40
+260 / +500
°C / °F
temp_coef1
2.1
0
+10 %/°C
temp_coef1
2.1
0
+10 %/°C
cell_constant_factory
0.1 0.005 50 cm-1
USP_safety_margin
0
0 +100 %
Measurement Warning limits
temp_warning_high_limit
+260 / +500
-40 / -40
+260 / +500
°C / °F
temp_warning_low_limit
-40 / -40
-40 / -40
+260 / +500
°C / °F
conductance_warning_high_limit 0.25
0
500
S
conductance_warning_low_limit 1
0
500
S
resistance_warning_high_limit
2000000
0
100000000
Ω
resistance_warning_low_limit
0.2
0
100000000
Ω
polarization_high_limit 15 0 50
%
Calibration limits
c.c._high_limit
120 100 120 %
c.c._low_limit
80 80 120 %
Calibration settings
stabilization_time
5
2
30 seconds
step_range
1 0.1 20 %
calibration_interval
250 1 250
days
Communication settings
Modbus_address
1 1 247
Serial_profile
1 1 2
Diagnostics settings
diag_progress_time_bad_limit 2000
1
10000 days
diag_heat_cycle_bad_limit
500
10
1000 cycles
diag_heat_cycle_temp
+50 / +122
+1 / +33.8
+250 / +482
°C / °F
diag_heat_cycle_time
10 0.1 60
minutes
sterilized_temp
+250/ +482
0 / +32
+250/ +482
°C / °F
sterilized_time
100 0 100
minutes
high_temp1
+250/ +482
-20 / -4
+250/ +482
°C / °F
high_temp2
+250/ +482
-20 / -4
+250/ +482
°C / °F
low_temp
-20 / -4
-20 / -4
+250/ +482
°C / °F
NaN: Not a Number