Item
Factory Setting limits Unit
default
Lower
Upper
limit limit
Calibration limits
zero_high_limit
+120 0 +600 mV
zero_low_limit
-120 -600 0 mV
slope_high_limit
+110 +100 +110 %
slope_low_limit
+70 +70 +100 %
ORP_zero_high_limit
+120 0 +500 mV
ORP_zero_low_limit
-120 -500 0 mV
ORP_slope_high_limit
+110 +100 +110 %
ORP_slope_low_limit
+70 +70 +100 %
Calibration settings
stabilization_time
5
2
30 seconds
pH_step_range
+0.03 +0.01 +1.00 pH
ORP_step_range
+1 +1 +100 mV
rH_step_range
0.05 0.01 10.00
calibration_interval
250 1 250
days
buffer_acceptance_range
1.000 0.001 5.000 pH
Communication settings
Modbus_address
1 1 247
Serial_profile
1 1 2
Diagnostics settings
diag_imp_pHORP_fine
10
1
10
MΩ
diag_imp_ref_fine
10
1
10
MΩ
diag_progress_time_bad_limit
2000 1 10000 days
diag_heat_cycle_bad_limit
500 10 1000 cycles
diag_heat_cycle_temp
+50 / +122 +1 / +33.8 +250 / +482
°C / °F
diag_heat_cycle_time
10 0.1 60
minutes
sterilized_temp
+140 / +284 0 / +32 +140 / +284
°C / °F
sterilized_time
100 0 100
minutes
high_temp1
+140 / +284 -30 / -22 +140 / +284
°C / °F
high_temp2
+140 / +284 -30 / -22 +140 / +284
°C / °F
low_temp
-30 / -22
-30 / -22 +140 / +284
°C / °F
high_pH_value
+13 -2 +16 pH
low_pH_value
+1 -2 +16 pH
NaN: Not a Number