11 APPENDICES
11.1 Appendix 1: Factory defaults & setting limits pH
The following table shows the details of the factory defaults of the SA11 and their setting
limits. Setting can be done by HOST system but some setting limits are restricted by
limitation of HOST.
Table A1.1: Factory defaults & setting limits pH
Item
Factory Setting limits Unit
default
Lower
Upper
limit limit
Measurement settings
manual_temp
+25 / +77
-40 / -40 +260 / +500
°C / °F
temp_ref
+25 / +77
-40 / -40 +260 / +500
°C / °F
temp_offset
0
-10 / +14
+10 / +50
°C / °F
external_temp
NaN
-40 / -40 +260 / +500
°C / °F
pH_temp_coef
0
-0.1
+0.1
pH/ºC or
pH/ºF
ORP1_temp_coef
0
-10
+10
mV/ºC or
ORP2_temp_coef
0 -10 +10 mV/ºF
pH_grab_offset
0 -300 +300 mV
pH_zero
0 -600 +600 mV
pH_slope
+100 +70 +110 %
pH_ITP
+7 0 +14 pH
pH_zero2
NaN -600 +600 mV
pH_slope2
NaN +70 +110 %
ORP1_grab_offset
0 -500 +500 mV
ORP1_zero
0 -500 +500 mV
ORP1_slope
+100 +70 +110 %
ORP2_grab_offset
0 -500 +500 mV
ORP2_zero
0 -500 +500 mV
ORP2_slope
+100 +70 +110 %
Measurement Warning limits
temp_warning_high_limit
+260 / +500 -40 / -40 +260 / +500
°C / °F
temp_warning_low_limit
-40 / -40
-40 / -40 +260 / +500
°C / °F
pH_warning_high_limit
(note)
+16 -2 +16 pH
pH_warning_low_limit (note)
-2
-2
+16
pH
ORP_warning_high_limit
+1500 -1500 +1500 mV
ORP_warning_low_limit
-1500 -1500 +1500 mV
rH_warning_high_limit
100 0 100
rH_warning_low_limit
0 0 100
Imp_pHORP low_warning_high_limit
200
1
+1000
kΩ
Imp_pHORP low_warning_low_limit
1
1
+1000
kΩ
Imp_ref low_warning_high_limit
200
1
+1000
kΩ
Imp_ref low_warning_low_limit
1
1
+1000
kΩ