IM 12A06S01-00EN-P
27
Table 4.3: Measurement warnings on calculated outputs
Failure too high means that the calculated output value is higher than the user configured
limit. Failure too low means that the calculated output value is less than the user configured
limit.
Warning item
Warning based upon conditions set
Temp. too high
Limit, to be configured by user from
-40°C / -40°F or -30°C / -22°F up to
+260°C / +500°F or +140°C / +284°F
Temp. too low
Limit, to be configured up to
-40°C / -40°F or -30°C / -22°F up to
+260°C / +500°F or +140°C / +284°F
pH too high
Limit, to be configured by user from
-2 pH up to 16 pH
pH too low
Limit, to be configured by user from
-2 pH up to 16 pH
ORP too high (note 1)
Limit, to be configured by user from
-1500 mV up to +1500 mV
ORP too low
Limit, to be configured by user from
-1500 mV up to +1500 mV
rH too high
Limit, to be configured by user from
0 up to 100
rH too low
Limit, to be configured by user from
0 up to 100
IMP1 too high (note 2)
Limit, to be configured by user from
1 (kΩ) up to 1000 (kΩ)
IMP1 too low (note 2)
Limit, to be configured by user from
1 (kΩ) up to 1000 (kΩ)
IMP2 too high (note 2)
Limit, to be configured by user from
1 (kΩ) up to 1000 (kΩ)
IMP2 too low (note 2)
Limit, to be configured by user from
1 (kΩ) up to 1000 (kΩ)
pH temp. comp.
The temperature compensation is result
of extrapolation of the matrix values (see
section 4.1.4)
SA11 internal temp. outside operating limits The internal temperature is outside the
operating limits from -30°C up to +125°C
Note 1:
ORP can mean ORP1 or ORP2, depends on ORP setting (see section 4.1.6).
Note 2:
IMP1 and/or IMP2 measurement warning only active if IMP1/2 are configured as
low.