Waveform Analysis
Measurement Range (Time Range)
As the default, measurement starts at -4div from the display center and extends to +4div, but
you can shorten the range. You make the setting by positioning two vertical lines: a dot-dash
line for the start point, and a dashed line for the end point.
Measurement parameters (Items)
You can select from among the following 26 parameters. You can measure up to a total of
sixteen parameters simultaneously.
• Voltage-axis parameters
P-P
Max
Min
High
Low
+Overshoot
-Overshoot
Max
: Maximum voltage [V]
Avg
: Average voltage (1/n)
Σ
xi [V]
Min
: Minimum voltage[V]
Rms
: Root mean squeare value (1/ n )(
Σ
(xi)2)1/2[V]
High
: High level voltage [V]
P-P
: Peak to peak value (Max-Min) [V]
Low
: Low level voltage[V]
StdDev : Standard deviation (1/n(
Σ
xi2-
Σ
(xi)2)1/2
Amp
: Amplitude (High-Low) [V]
+OShot : Overshoot (Max-High)/(High-Low) [%]
-OShot : Overshoot (Low-Min)/(High-Low) [%]
• Time-axis parameters
+Duty
: +Duty rate (+Width)/Period
×
100 [%]
-Duty
: -Duty rate (–Width)/Period
×
100 [%]
Freq
: Frequency (Hz]
Period
: Period [s]
Rise
: Rise time [s]
Fall
: Fall time [s]
+Width
: Width above mesial value [s]
-Width
: Width below mesial value [s]
+ Width
- Width
Period
Distal line
(90%)
Mesial line(50%)
Proximal line
(10%)
Rise
Fall
RDelay,FDelay
: Delay time [s]
CH1
CH2
CH1
CH2
RDelay
FDelay
RDelay
CH2 RDelay/FDelay(When Delay Ref = CH1)
FDelay
Delay Ref line
Delay Ref line
9.2 Automatic Measurement of Waveform Parameters