7-4
IM 701830-01E
7.3
Sequential Store Mode
≡
For a description of this function, refer to page 1-13.
≡
Function
Acquisition count
Available numerical settings are as follows.
If you set the value to Infinite, acquisition will continue until you switch it off with the
START/
STOP
key.
Infinite, 2 to 1000
However, the count number that can be selected varies depending on the selected record length
and acquisition mode.
For details, refer to Appendix 2 “Relationship between the Record Length and Acquisition
Mode.”
Waveform display method
You can recall waveforms from memory in the same way as you do when working with the
history function. For details, refer to Section 7.5, “Using the History Memory.”
Restrictions and precautions
• You can not use this mode together with the roll mode.
• If you stop waveform acquisition by pressing the
START/STOP
key, sequential storage also
stops. It then restarts again from the beginning when acquisition resumes.
• The waveform is acquired in normal mode regardless of the trigger mode setting. If no
trigger occurs, sequential store does not finish.
• If the acquisition count is set to “Infinite,” it operates in the same way as the normal
waveform acquisition (normal mode).
Operating Procedure
Set the acquisition mode to sequential store
1.
Press the
ACQ
key.
2.
Press the “
Acq Mode
” soft key to open the acquisition-mode menu.
3.
Press the “
Acq Mode
” soft key as necessary to select Sequence.
Set the acquisition count
4.
Press the “
Count
” soft key, so that the jog shuttle action is set to “
Count
.”
5.
Set the count by turning the jog shuttle.