8-7
8
IM 701830-01E
Display
8.5
Turning the Extra Window ON/OFF
≡
For a description of this function, refer to page 1-15.
≡
Function
An extra window is provided to display information other than waveforms when waveforms
obscure the information in the display area. The following three types of information can be
displayed in the extra window.
Cursor measure values
Automatically measured wave form parameter
Digital values of measuring data
Cursor measured values
Extra Window
The number of display points available in the vertical direction when the extra
window is ON
When the extra window is set to ON, the vertical size of the waveform display area is reduced
by 2/3. However, the vertical resolution remains unchanged.
ON : 288 points
OFF : 432 points
(When the display format is “Single”)
Note
Use of the extra window together with certain zoom formats may cause scale values (at the top and bottom
of each waveform window) to overlap, so that they become difficult to read.
Operating Procedure
1.
Press the
DISPLAY
key.
2.
Press the “
Extra Window
” soft key to select “
ON
” or “
OFF
.”