6-8
IM 701830-01E
6.7
Setting the Trigger Delay
≡
For a description of this function, refer to page 1-11.
≡
Function
Although the display usually shows the waveform before and after the trigger point, using the
delay function, it is possible to display the acquired waveform after a fixed time period elapses.
Setting range for trigger delay
0 to 1 s (Resolution is 500 ns)
Delay time
Trigger point
Trigger position
Points for attention
When T/div is changed, the trigger delay remains unchanged.
Operating Procedure
1.
Press the
POSITION/DELAY
key.
2.
Press the “
Delay
” soft key to set jog shuttle action to Delay.
3.
Turn the jog shuttle to set the desired trigger delay.
You can change the setting a digit using the arrow keys (located below the jog shuttle).
You can reset the delay to 0s by pressing the
RESET
key.