Operating Procedure
Setting the trigger type
1.
Press the
ENHANCED
key.
2.
Press the “
Type
” soft key to display the trigger type menu.
3.
Keep pressing the “
Type
” soft key to select “
B>Time
,” “
B<Time
,” or “
B Time Out
.”
Setting the channel status for condition B
4.
Press the “
Set Pattern
” soft key to display the pulse width trigger setting menu.
5.
Turn the jog shuttle to move the cursor to “
CH1
” under <Parallel B>.
6.
Press the
SELECT
key to display the <Parallel A> CH1 menu.
7.
Press the
SELECT
key again to select “
H
,” “
L
,” or “
X
.”
8.
Repeat the above steps to set the channel status for CH2 to CH16.
Setting the pulse width
9.
Turn the jog shuttle to move the cursor to the “
Time
”.
10.
Press the
SELECT
key to display the “
Time
” menu.
11.
Turn the jog shuttle to set the desired pulse width.
You can change the value of the digit using the cursor key (located below the jog shuttle).
Pressing the
SHIFT
key, then turning the jog shuttle resets the pulse width to 0.2
µ
s.
Other settings
12.
You must also set the following parameters: trigger mode, trigger level, trigger hysteresis,
and holdoff time. For information about how to set these parameters, refer to Sections 6.1 to
6.7.
Note
Status settings for condition B apply in common to all trigger types.
6.13 Setting the B>Time, B<Time or B Time Out Trigger (ENHANCED)