22
4. OPERATION PROCEDURES
Auto Stop Mode
Auto stop operation
procedure
This mode is used to specify when unit should terminate constant
temperature operation. In contrast to quick auto stop mode, this
mode must be set before operation.
1. Set stop time
①
Press the TIMER key from the initial settings screen.
②
Mode used in the previous session will be shown in the
center display. Press the TIMER key again and center
display will begin flashing. Use TIMER key to select
, signifying auto stop operation, then press the
ENTER key.
Top display will show
, indicating that temperature
can be set. Current temperature setting flashes in center
display. AUTO STOP lamp also flashes.
③
Set temperature using
▼▲ keys.
④
Press the ENTER key again.
Top display will show
, indicating that timer can be
set. Current timer setting will flash in center display.
⑤
Set timer using the
▼▲ keys.
Timer function:
Maximum setting for timer is "999 hours and 50 minutes ".
The time can be set in increments of one minute, under 99 hours
and 59 minutes.
Setting increments are 10 minutes, after 100 hours.
When the
▼▲ keys are held down, values advance perpetually.
Press repeatedly for incremental adjustment.
2. Start auto stop operation
Press the RUN/STOP key for about one second after setting
timer.
AUTO STOP and HEATER lamp illuminate, indicating auto stop
operation mode has started.
Timer begins counting down when chamber temperature
reaches objective temperature.
①
②
④
③