16
4. OPERATION PROCEDURES
Operation Modes and Functions
Operation modes for this unit are defined in the table below:
N
o.
Name
Description
Page
1
Constant Temperature
Mode
Pressing the FIXED TEMP key brings up constant
temperature setup mode.
The "▲▼" keys are used to set temperature.
Pressing the RUN/STOP key initiates or terminates
operation.
20
2
Quick Auto Stop
Mode
This mode is used to automatically terminate an operation
when a specified time period has passed (decided during
operation).
This is done by pressing the TIMER key at any time during
constant temperature operation.
The "▲▼" keys are used to set the timer.
Pressing the RUN/STOP key begins operation in quick auto
stop mode, activates the timer (once temp setting is
reached) and automatically terminates operation when
timer reaches 0:00.
21
3
Auto Stop Mode
This mode is used to automatically terminate an operation
when a specified time period has passed (decided before
operation).
Pressing the TIMER key displays "AStp".
The temperature (Sv) is set by pressing the ENTER key.
The operation time (tim) is set by pressing ENTER again.
Pressing the RUN/STOP key begins operation in auto stop
mode. Operation automatically terminates when timer
reaches 0:00.
22
4
Auto Start Mode
This operation is used to automatically begin an operation
after a specified period has passed.
Pressing the TIMER key displays "AStr".
The temperature (SV) is set by pressing the ENTER key.
The operation time (tim) is set by pressing ENTER again.
Pressing the RUN/STOP key begins operation in auto start
mode.
24
5
Programmed Operation
This operation is used to run a combination of modes, times
and temperatures as one operation.
Pressing the PROGRAM key displays "PrG1".
Press it again to select program mode.
Press the ENTER key to select the pattern (Pat). Press the
ENTER key to display "End". Enter the number of patterns
to be used.
Enter step temperatures and step times for "Sv_x" and "t_x"
respectively (where
“x” stands for a number or value).
26
*
It is not possible to change modes during operation. If a mode change is required, operation must
be terminated and a new mode of operation must be set.