27
4. OPERATION PROCEDURES
Building Programs
Temperature fall/rise
curve for DKN-C
models
Temperature rise and fall curves for DKN-C models are shown below in
50
C increments.
Numeric values signify time needed (in minutes) for temperature to rise
or fall for every 50
C. (example: DKN-C model 602 needs
approximately 15 to increase from 100
C to 150
C) Temperature
stabilization time is an added factor and not included in the table below.
Be sure to conduct a test run before finalizing program pattern times.
Conditions: room temperature 23
C, no load (unit: minute)
Building Programs
The program pattern below will be used as an example for bulding programs
1. Turn on main circuit breaker (MCB ON
“|”)
Initial values will be shown for about 4 seconds after power-on,
then displays will switch to the initial settings screen, showing
current chamber temperature (top), operation mode character
(center) and overheat prevention setting (bottom).
2. Select program mode/program pattern
Press the PROGRAM key once.
The top display will show previously used program.
Pressing the PROGRAM key again will cause program mode to
begin flashing.
Pressing the PROGRAM key repeatedly thereafter will toggle
between the 3 available program modes (while flashing).
DKN302C
DKN402C/412C
DKN602C/612C
DKN812C
DKN912C
Rise
Fall
Rise
Fall
Rise
Fall
Rise
Fall
Rise
Fall
210
℃
5
-
5
-
5
-
5
-
5
-
200
℃
15
5
20
5
20
5
15
5
25
5
150
℃
10
15
15
20
15
20
10
25
20
30
100
℃
10
20
15
30
15
30
10
40
20
60
50
℃
5
45
5
60
5
75
5
120
5
180