24
4. OPERATION PROCEDURES
Auto Start Mode
Auto start operation
procedure
This mode is used to specify an automatic start time for constant
temperature operation.
1. Set start time
①
Press the TIMER key from the initial settings screen.
②
Previous operation mode is displayed in center display.
Press the TIMER key again and center display begins
flashing. Use TIMER key to select
, signifying
auto start mode, then press the ENTER key.
Top display will show
, indicating temperature
can now be set. Current temperature setting flashes in the
center display. AUTO START lamp also flashes.
③
Set temperature using the
▼▲ keys.
④
Press the ENTER key again.
Top display shows
, indicating the timer can be set.
Current timer setting will flash in the center display.
⑤
Set timer using the
▼▲ keys.
Timer function:
Maximum timer setting is "999 hours and 50 minutes ".
The time can be set in increments of one minute, under 99 hours
and 59 minutes.
Setting increments are 10 minutes, after 100 hours.
When
▼▲ keys are held down, values advance perpetually.
Press repeatedly for incremental adjustment.
2. Initialize auto start operation
Press the RUN/STOP key for about one second after setting
timer.
Auto start operation mode begins when timer reaches 0.00 and
AUTO START lamp illuminates.
①
②
③
④