background image

34

35

Silver Anode
After extended use, a thick layer of Silver Chloride (AgCl) builds up 
on the silver anode (figure 11) reducing the sensitivity of the sensor. 
The anode must be cleaned to remove this layer and restore proper 
performance. The cleaning can be chemical and/or mechanical:

Mechanical cleaning: In order to sand the silver anode, simply hold the 
sensor in a vertical position. Wet the 400 grit sanding disc and gently 
wrap it around the sensor and twist it a few times to lightly sand the 
anode (the goal is to simply sand off any build-up without scratching 
or removing layers of the anode itself). Usually, 3 to 4 twists of the 
sanding disc are sufficient to remove deposits. 

After completing the sanding procedure, repeatedly rinse the 
electrode with distilled or deionized water and wipe with lens cleaning 
tissue to remove any grit left by the sanding disc. Thoroughly rinse the 
entire tip of the sensor with distilled or deionized water and install a 
new membrane.

Chemical cleaning: Remove the cap membrane and rinse the sensor 
with deionized or distilled water. Soak the sensing section of the sensor 
in a 14% ammonium hydroxide solution for 2 to 3 minutes or in a 3% 
ammonia solution overnight for 8-12 hours (most household ammonia 
cleaners are typically around 3%). Rinse heavily in cool tap water 
followed by a thorough rinsing with distilled or deionized water. The 
anode should then be thoroughly wiped with a lens cleaning tissue to 
remove the residual layer from the anode. Trapping residual ammonia 
under the new membrane cap can quickly tarnish the electrode and/or 
give false readings and should therefore be avoided. 

After performing a chemical cleaning on the polarographic sensor, 
lightly sand the cathode and anode following the mechanical cleaning 
procedures described previously.

If this procedure is unsuccessful, as indicated by improper sensor 
performance, contact YSI Technical Support or the YSI authorized 
dealer where you purchased the instrument.

Galvanic Sensors – Model # 605202

YSI recommends that the Sodium Chloride (NaCl) electrolyte solution 
and cap membrane be changed every 2-4 weeks during regular use. 
In addition, the electrolyte solution and membrane should be changed 
if (a) bubbles are visible under the membrane; (b) dried electrolyte 

is visible around the membrane; and (c) if the sensor shows unstable 
readings or other sensor-related symptoms.

The Galvanic dissolved oxygen sensor is continuously reducing oxygen 
even when the instrument is turned off. This factor allows the sensor to 
be used with no warm-up period as soon as the instrument is powered 
on. However, because the sensor is “on” all the time, some solid from 
the oxidation of the zinc anode will form in the electrolyte within 1-2 
weeks of activation. Small amounts of the solid will generally cause 
no performance problems, but excessive amounts may result in jumpy 
dissolved oxygen readings. The rate of solid formation is dependent 
on the type of membrane installed. The formation of solids typically 
form more rapidly with 5913 (1.25 mil PE) membrane, and less rapid 
with 5914 (2 mil PE) membranes.

At the time the cap membrane is changed, YSI recommends that you 
rinse the anode and cathode (figure 11) with distilled or deionized 
water and wipe with a lens cleaning tissue. If white deposits are evident 
on the anode after cleaning, YSI recommends that you remove this 
material by sanding the anode with the 400 grit wet/dry sanding disc 
included in the membrane kit following the “Mechanical Cleaning” 
instructions under the Polarographic Silver Anode maintenance section 
of this manual.  If there are deposits on the cathode, sand the cathode 
with the 400 grit wet/dry sanding disc following the maintenance 
instructions listed in this manual for the Polarographic Gold Cathode.  

If this procedure is unsuccessful, as indicated by improper sensor 
performance, contact YSI Technical Support or the YSI authorized 
dealer where you purchased the instrument.

SENSOR STORAGE

SHORT TERM STORAGE

The instrument is supplied with a plastic calibration/storage cup that 
screws onto the cable. The cup is used for short-term storage (less than 
1 weeks). Be sure to keep a small amount of clean tap water in the cup 
during storage. With a small amount of water, the cup provides a 100% 
water saturated air environment which is ideal for short-term sensor 
storage. 

LONG TERM STORAGE  

The dissolved oxygen sensor should be stored long term in a dry state 
while the pH or ORP sensor should be stored in solution. When storing 
for more then 30 days, place the pH or ORP sensor in the storage 

Содержание YSI Pro1020

Страница 1: ...USER MANUAL English Fran ais Espa ol Portugu s Pro1020 User Manual DOCUMENT 605187...

Страница 2: ...1 Battery Installation 1 Key Pad 2 Connecting the Sensor and Cable 3 Membrane Installation 6 Run Screen 7 Backlight 7 Powering Off 7 Navigation 7 First Power On 8 System Setup Menu 9 DO Local 10 Last...

Страница 3: ...YSI will repair or replace at its sole discretion free of charge any product that YSI determines to be covered by this warranty To exercise this warranty call your local YSI representative or contact...

Страница 4: ...arts or materials are damaged or missing contact YSI Customer Service at 800 897 4151 1 937 767 7241 or the authorized YSI distributor from whom the instrument was purchased BATTERY INSTALLATION The i...

Страница 5: ...he bottom of the Run screen and to increase numerical inputs 3 Power and Backlight Press once to turn instrument on Press a second time to turn backlight on Press a third time to turn backlight off Pr...

Страница 6: ...Next remove the o ring and slide the lid off the sensor INSTALLING THE DISSOLVED OXYGEN SENSOR Before installing a sensor or connecting the cable to the instrument the Sensor Type must be configured f...

Страница 7: ...cap from the dissolved oxygen sensor by pulling it straight off Thoroughly rinse the sensor tip with distilled or deionized water Fill the cap membrane 3 4 full of electrolyte solution then tap the ca...

Страница 8: ...dissolved oxygen sensor type and membrane type Use the up or down arrow keys to highlight the appropriate language sensor and membrane then press enter to confirm figure 6 The Sensor Type must be con...

Страница 9: ...instead of during calibration Example Instrument 1 with DO Local enabled At 737 mmHg barometric pressure roughly 841 ft above sea level the instrument would calibrate to 100 When taking measurements...

Страница 10: ...ompatible sensors for use with a field cable Polarographic This sensor has a black sensor body and is engraved with the model number 2003 Polarographic will be abbreviated Polaro in the instrument men...

Страница 11: ...lay steadily and the instrument will beep twice if Audio is turned on DO Auto Stable can be set to a change of 0 0 to 1 9 over 3 to 19 seconds ISE Auto Stable can be set to a change of 0 0 to 9 9 over...

Страница 12: ...nge the language Highlight the desired language English Spanish Portuguese or French and press enter to enable The enabled language will have an X in the box next to it Highlight the ESC Exit box and...

Страница 13: ...ate formula for this uncorrection is True BP Corrected BP 2 5 Local Altitude in ft above sea level 100 Although the Pro1020 barometer range is 400 0 to 999 9 mmHg you can not adjust the value across t...

Страница 14: ...ubleshooting guide for possible solutions CALIBRATING IN PERCENT DO LOCAL ENABLED Perform this calibration procedure when DO Local is enabled and Quick DO Cal is disabled in the System Setup menu 1 Pe...

Страница 15: ...5 If necessary use the up and down arrow keys to adjust the pH buffer value Note the pH mV reading which ideally should be between 50 and 50 in buffer 7 6 Press enter to continue to second point 7 Ri...

Страница 16: ...membrane If using the 1001A amplified pH sensor be sure to use the guard extender supplied in the 605216 amplified pH sensor kit The extender is screwed onto the sensor bulkhead before the guard or c...

Страница 17: ...enter to regain access to scrolling through the saved data The data set displayed is indicated by the data set number 01 through 50 ERASING DATA While viewing saved data press the enter key to access...

Страница 18: ...or connector Check the o ring and the groove for any excess grease or contamination If contamination is evident clean the o ring and nearby plastic parts with lens cleaning tissue or equivalent lint f...

Страница 19: ...the sensor in clean water wipe with a cotton swab saturated with clean water not DI water and then rerinse with clean water To be certain that all traces of the acid are removed i from the sensor cre...

Страница 20: ...d in the membrane kit Do not sand the electrodes every membrane change as this is not routine maintenance In fact visually the anode may appear tarnished and operate properly YSI recommends using the...

Страница 21: ...embrane should be changed if a bubbles are visible under the membrane b dried electrolyte is visible around the membrane and c if the sensor shows unstable readings or other sensor related symptoms Th...

Страница 22: ...em for service Symptom Possible Solution Instrument will not calibrate pH or ORP instrument displays Calibration Over Calibration Under or Unstable Reading during calibration 1 Verify correct ISE sens...

Страница 23: ...specifications represent typical performance and are subject to change without notice For the latest product specification information please visit YSI s website at ysi com or contact YSI Tech Suppor...

Страница 24: ...E blue cap 6 each includes electrolyte Used with Galvanic sensor All cables include a temperature sensor The pH or ORP and Dissolved Oxygen sensors are sold separately D E C L A R AT I O N O F C O N F...

Страница 25: ...country and region and users are expected to understand and follow the battery disposal requirements for their specific locale C O N TA C T I N F O R M AT I O N ORDERING AND TECHNICAL SUPPORT Telephon...

Страница 26: ...Item 605187 Rev A January 2014 2014 YSI Incorporated...

Отзывы: