background image

18

19

Parameter

Reset Defaults

ISE Sensor Type

pH

ISE Units

pH

ISE Auto Stable

Off (0.0 % Change 

and 10 seconds)

pH Buffer Set

USA

Temperature Units

°C

Pressure Units

mmHg

Language

English

Auto Shutoff

30 minutes

Dissolved Oxygen Calibration

Reset to 100% for enabled 

membrane and sensor*

ISE Calibration

Reset to factory default*

Barometer Calibration

Reset to factory default*

*It is recommended to perform a barometer, dissolved oxygen and ISE  
calibration after performing a reset.

C A L I B R A T I O N 

TEMPERATURE 

All Pro1020 cables have built-in temperature sensors. Temperature 
calibration is not required nor is it available.

BAROMETER

The barometer in the Pro1020 is calibrated at the factory. The barometer 
reading must be accurate to ensure accurate DO% calibrations and 
readings. If the  barometer requires an adjustment, use the up or down 
arrow keys to highlight the barometer box along the bottom of the 
Run screen, then press enter. Next, use the up or down arrow keys to 
adjust the barometer reading to the local, true barometric pressure. 
Continually depress the up or down arrow key to change the barometer 
value more rapidly. Press enter to confirm and save the barometer 
adjustment.  

Do not use a barometer value that is corrected to sea level. 
Laboratory barometer readings are usually “true” (uncorrected) 
values of air pressure and can be used “as is” for barometer 
calibration. Weather service readings are usually not “true”, i.e., 
they are corrected to sea level, and therefore cannot be used 
until they are “uncorrected”. An approximate formula for this 
“uncorrection” is:

True BP = [Corrected BP] – [2.5 * (Local Altitude in ft above sea level/100)]

Although the Pro1020 barometer range is 400.0 to 999.9 mmHg, you 
can not adjust the value across the entire range. The barometer is 
very accurate and the instrument will not allow you to adjust the value 
drastically beyond what it is measuring.    

DISSOLVED OXYGEN

The dissolved oxygen sensor can be easily calibrated with the press 
of two keys by enabling Quick DO Cal in the System Setup menu and 
following the Quick DO Calibration procedure.
  
Ensure the barometer is reading accurately before performing a Quick 
DO Cal, DO% or DO Local% calibration because these calibration 
procedures use the barometer reading during calibration. If the 
barometer reading is erroneous during a calibration, the dissolved 
oxygen measurements will be inaccurate.   

QUICK DO CALIBRATION

Perform this calibration procedure when Quick DO Calibration is 
enabled in the System Setup menu.  

1.  Ensure the DO sensor has a good membrane with electrolyte 

installed. A good membrane is free of wrinkles, tears, fouling 
and air bubbles. Install the sensor guard onto the probe.  

2.  Place a small amount of clean water in the calibration/storage 

cup and screw it onto the bulkhead. Do not thread it on 
completely. Be sure that 2 or 3 threads are disengaged to 
ensure the cup is vented to the atmosphere. Additionally, there 
should only be little bit of water at the bottom of the calibration 
cup. There should not be water droplets or water touching the 
DO membrane or temperature sensor. 

3.  Power the instrument on and, if using a Polarographic sensor, 

wait approximately 5 to 15 minutes for the storage cup to 
become completely saturated and for the sensor to stabilize. 

i

Содержание YSI Pro1020

Страница 1: ...USER MANUAL English Fran ais Espa ol Portugu s Pro1020 User Manual DOCUMENT 605187...

Страница 2: ...1 Battery Installation 1 Key Pad 2 Connecting the Sensor and Cable 3 Membrane Installation 6 Run Screen 7 Backlight 7 Powering Off 7 Navigation 7 First Power On 8 System Setup Menu 9 DO Local 10 Last...

Страница 3: ...YSI will repair or replace at its sole discretion free of charge any product that YSI determines to be covered by this warranty To exercise this warranty call your local YSI representative or contact...

Страница 4: ...arts or materials are damaged or missing contact YSI Customer Service at 800 897 4151 1 937 767 7241 or the authorized YSI distributor from whom the instrument was purchased BATTERY INSTALLATION The i...

Страница 5: ...he bottom of the Run screen and to increase numerical inputs 3 Power and Backlight Press once to turn instrument on Press a second time to turn backlight on Press a third time to turn backlight off Pr...

Страница 6: ...Next remove the o ring and slide the lid off the sensor INSTALLING THE DISSOLVED OXYGEN SENSOR Before installing a sensor or connecting the cable to the instrument the Sensor Type must be configured f...

Страница 7: ...cap from the dissolved oxygen sensor by pulling it straight off Thoroughly rinse the sensor tip with distilled or deionized water Fill the cap membrane 3 4 full of electrolyte solution then tap the ca...

Страница 8: ...dissolved oxygen sensor type and membrane type Use the up or down arrow keys to highlight the appropriate language sensor and membrane then press enter to confirm figure 6 The Sensor Type must be con...

Страница 9: ...instead of during calibration Example Instrument 1 with DO Local enabled At 737 mmHg barometric pressure roughly 841 ft above sea level the instrument would calibrate to 100 When taking measurements...

Страница 10: ...ompatible sensors for use with a field cable Polarographic This sensor has a black sensor body and is engraved with the model number 2003 Polarographic will be abbreviated Polaro in the instrument men...

Страница 11: ...lay steadily and the instrument will beep twice if Audio is turned on DO Auto Stable can be set to a change of 0 0 to 1 9 over 3 to 19 seconds ISE Auto Stable can be set to a change of 0 0 to 9 9 over...

Страница 12: ...nge the language Highlight the desired language English Spanish Portuguese or French and press enter to enable The enabled language will have an X in the box next to it Highlight the ESC Exit box and...

Страница 13: ...ate formula for this uncorrection is True BP Corrected BP 2 5 Local Altitude in ft above sea level 100 Although the Pro1020 barometer range is 400 0 to 999 9 mmHg you can not adjust the value across t...

Страница 14: ...ubleshooting guide for possible solutions CALIBRATING IN PERCENT DO LOCAL ENABLED Perform this calibration procedure when DO Local is enabled and Quick DO Cal is disabled in the System Setup menu 1 Pe...

Страница 15: ...5 If necessary use the up and down arrow keys to adjust the pH buffer value Note the pH mV reading which ideally should be between 50 and 50 in buffer 7 6 Press enter to continue to second point 7 Ri...

Страница 16: ...membrane If using the 1001A amplified pH sensor be sure to use the guard extender supplied in the 605216 amplified pH sensor kit The extender is screwed onto the sensor bulkhead before the guard or c...

Страница 17: ...enter to regain access to scrolling through the saved data The data set displayed is indicated by the data set number 01 through 50 ERASING DATA While viewing saved data press the enter key to access...

Страница 18: ...or connector Check the o ring and the groove for any excess grease or contamination If contamination is evident clean the o ring and nearby plastic parts with lens cleaning tissue or equivalent lint f...

Страница 19: ...the sensor in clean water wipe with a cotton swab saturated with clean water not DI water and then rerinse with clean water To be certain that all traces of the acid are removed i from the sensor cre...

Страница 20: ...d in the membrane kit Do not sand the electrodes every membrane change as this is not routine maintenance In fact visually the anode may appear tarnished and operate properly YSI recommends using the...

Страница 21: ...embrane should be changed if a bubbles are visible under the membrane b dried electrolyte is visible around the membrane and c if the sensor shows unstable readings or other sensor related symptoms Th...

Страница 22: ...em for service Symptom Possible Solution Instrument will not calibrate pH or ORP instrument displays Calibration Over Calibration Under or Unstable Reading during calibration 1 Verify correct ISE sens...

Страница 23: ...specifications represent typical performance and are subject to change without notice For the latest product specification information please visit YSI s website at ysi com or contact YSI Tech Suppor...

Страница 24: ...E blue cap 6 each includes electrolyte Used with Galvanic sensor All cables include a temperature sensor The pH or ORP and Dissolved Oxygen sensors are sold separately D E C L A R AT I O N O F C O N F...

Страница 25: ...country and region and users are expected to understand and follow the battery disposal requirements for their specific locale C O N TA C T I N F O R M AT I O N ORDERING AND TECHNICAL SUPPORT Telephon...

Страница 26: ...Item 605187 Rev A January 2014 2014 YSI Incorporated...

Отзывы: