background image

16

17

mVs can help you determine if you are performing a good calibration 
and the condition of the pH electrode. 

When ORP is enabled in the ISE Sensor Type menu, only ORP mVs can 
be enabled as the ISE unit.  

pH BUFFER SET

Highlight pH Buffer Set and press enter to open a submenu that allows 
you to select the Buffer Set used for auto buffer recognition during a 
pH calibration. There are two buffer set options: USA (4, 7 and 10) and 
NIST (4.01, 6.86 and 9.18). Highlight the buffer set and press enter to 
enable. The enabled buffer set will have an ‘X’ in the box next to it. 
Highlight the ESC-Exit box and press enter to save any changes and to 
close the submenu.

TEMPERATURE UNITS

Highlight Temperature Units and press enter to open a submenu that 
allows you to change the temperature units displayed on the Run 
screen. Highlight the desired unit (Celsius or Fahrenheit) and press 
enter to enable. The enabled temperature unit will have an ‘X’ in the 
box next to it. Only one unit may be enabled at a time. Highlight the 
ESC-Exit box and press enter to save any changes and to close the 
Temperature Units submenu.

PRESSURE UNITS

Highlight Pressure Units and press enter to open a submenu that 
allows you to change the barometric pressure units displayed on the 
Run screen. Highlight the desired unit (mmHg, inHg, mbar, psi, or kPa) 
and press enter to enable. The enabled pressure unit will have an ‘X’ 
in the box next to it. Only one unit may be enabled at a time. Highlight 
the ESC-Exit box and press enter to save any changes and to close the 
Pressure Units submenu.  

LANGUAGE

Highlight Language and press enter to open a submenu that allows 
you to change the language. Highlight the desired language (English, 
Spanish, Portuguese, or French) and press enter to enable. The enabled 
language will have an ‘X’ in the box next to it. Highlight the ESC-Exit 

box and press enter to save any changes and to close the Language 
submenu.

The text in the boxes along the bottom of the Run screen will always be 
displayed in English regardless of the language enabled in the System 
Setup menu.  

AUTO SHUTOFF

Auto Shutoff allows you to set the instrument to turn off automatically 
after a period of time after the last button press. In the setup menu, use 
the up or down arrow keys to highlight Auto Shutoff, then press enter 
to open the submenu. Press enter while the minute field is highlighted 
to make it adjustable. Next, use the up or down arrow keys to adjust the 
shut off time from 0 to 60 minutes. Press enter to save the new shutoff 
time. Next, highlight the ESC-Exit box and press enter to close the 
submenu. To disable Auto Shutoff, set the Time in Minutes to 0 (zero).

RESETTING THE SYSTEM SETUP MENU 

TO FACTORY DEFAULT

To reset the Pro1020 settings back to factory default, press the down 
arrow key while in the System Setup menu until the Reset -   box 
is highlighted, then press enter. The instrument will prompt you to 
confirm the reset. Highlight Yes and press enter to continue with the 
reset or highlight No and press enter to cancel the reset. A Factory 
Reset will not affect data saved in the instrument’s memory.  

The following will be set in the Pro1020 after performing a reset:

Parameter

Reset Defaults

DO Local%

Off

LDS (Last Digit Suppression)

Off

Quick DO Cal

Off

Audio

On

Contrast

Set to mid range

Dissolved Oxygen Sensor Type

Last Setting Confirmed

Dissolved Oxygen Membrane 
Type

Last Setting Confirmed

Dissolved Oxygen Units

mg/L and %

Dissolved Oxygen Auto Stable

Off (0.0 % Change 

and 10 seconds)

Содержание YSI Pro1020

Страница 1: ...USER MANUAL English Fran ais Espa ol Portugu s Pro1020 User Manual DOCUMENT 605187...

Страница 2: ...1 Battery Installation 1 Key Pad 2 Connecting the Sensor and Cable 3 Membrane Installation 6 Run Screen 7 Backlight 7 Powering Off 7 Navigation 7 First Power On 8 System Setup Menu 9 DO Local 10 Last...

Страница 3: ...YSI will repair or replace at its sole discretion free of charge any product that YSI determines to be covered by this warranty To exercise this warranty call your local YSI representative or contact...

Страница 4: ...arts or materials are damaged or missing contact YSI Customer Service at 800 897 4151 1 937 767 7241 or the authorized YSI distributor from whom the instrument was purchased BATTERY INSTALLATION The i...

Страница 5: ...he bottom of the Run screen and to increase numerical inputs 3 Power and Backlight Press once to turn instrument on Press a second time to turn backlight on Press a third time to turn backlight off Pr...

Страница 6: ...Next remove the o ring and slide the lid off the sensor INSTALLING THE DISSOLVED OXYGEN SENSOR Before installing a sensor or connecting the cable to the instrument the Sensor Type must be configured f...

Страница 7: ...cap from the dissolved oxygen sensor by pulling it straight off Thoroughly rinse the sensor tip with distilled or deionized water Fill the cap membrane 3 4 full of electrolyte solution then tap the ca...

Страница 8: ...dissolved oxygen sensor type and membrane type Use the up or down arrow keys to highlight the appropriate language sensor and membrane then press enter to confirm figure 6 The Sensor Type must be con...

Страница 9: ...instead of during calibration Example Instrument 1 with DO Local enabled At 737 mmHg barometric pressure roughly 841 ft above sea level the instrument would calibrate to 100 When taking measurements...

Страница 10: ...ompatible sensors for use with a field cable Polarographic This sensor has a black sensor body and is engraved with the model number 2003 Polarographic will be abbreviated Polaro in the instrument men...

Страница 11: ...lay steadily and the instrument will beep twice if Audio is turned on DO Auto Stable can be set to a change of 0 0 to 1 9 over 3 to 19 seconds ISE Auto Stable can be set to a change of 0 0 to 9 9 over...

Страница 12: ...nge the language Highlight the desired language English Spanish Portuguese or French and press enter to enable The enabled language will have an X in the box next to it Highlight the ESC Exit box and...

Страница 13: ...ate formula for this uncorrection is True BP Corrected BP 2 5 Local Altitude in ft above sea level 100 Although the Pro1020 barometer range is 400 0 to 999 9 mmHg you can not adjust the value across t...

Страница 14: ...ubleshooting guide for possible solutions CALIBRATING IN PERCENT DO LOCAL ENABLED Perform this calibration procedure when DO Local is enabled and Quick DO Cal is disabled in the System Setup menu 1 Pe...

Страница 15: ...5 If necessary use the up and down arrow keys to adjust the pH buffer value Note the pH mV reading which ideally should be between 50 and 50 in buffer 7 6 Press enter to continue to second point 7 Ri...

Страница 16: ...membrane If using the 1001A amplified pH sensor be sure to use the guard extender supplied in the 605216 amplified pH sensor kit The extender is screwed onto the sensor bulkhead before the guard or c...

Страница 17: ...enter to regain access to scrolling through the saved data The data set displayed is indicated by the data set number 01 through 50 ERASING DATA While viewing saved data press the enter key to access...

Страница 18: ...or connector Check the o ring and the groove for any excess grease or contamination If contamination is evident clean the o ring and nearby plastic parts with lens cleaning tissue or equivalent lint f...

Страница 19: ...the sensor in clean water wipe with a cotton swab saturated with clean water not DI water and then rerinse with clean water To be certain that all traces of the acid are removed i from the sensor cre...

Страница 20: ...d in the membrane kit Do not sand the electrodes every membrane change as this is not routine maintenance In fact visually the anode may appear tarnished and operate properly YSI recommends using the...

Страница 21: ...embrane should be changed if a bubbles are visible under the membrane b dried electrolyte is visible around the membrane and c if the sensor shows unstable readings or other sensor related symptoms Th...

Страница 22: ...em for service Symptom Possible Solution Instrument will not calibrate pH or ORP instrument displays Calibration Over Calibration Under or Unstable Reading during calibration 1 Verify correct ISE sens...

Страница 23: ...specifications represent typical performance and are subject to change without notice For the latest product specification information please visit YSI s website at ysi com or contact YSI Tech Suppor...

Страница 24: ...E blue cap 6 each includes electrolyte Used with Galvanic sensor All cables include a temperature sensor The pH or ORP and Dissolved Oxygen sensors are sold separately D E C L A R AT I O N O F C O N F...

Страница 25: ...country and region and users are expected to understand and follow the battery disposal requirements for their specific locale C O N TA C T I N F O R M AT I O N ORDERING AND TECHNICAL SUPPORT Telephon...

Страница 26: ...Item 605187 Rev A January 2014 2014 YSI Incorporated...

Отзывы: