background image

28

29

EXITING DATA MODE

While in Data mode, press enter to access the bottom functions. Next, 
highlight the ESC-Exit box and press enter to return to the Run screen.  

C A R E ,   M A I N T E N A N C E   A N D 
S T O R A G E

This section describes the proper procedures for care, maintenance 
and storage of the sensors. The goal is to maximize their lifetime and 
minimize down-time associated with improper sensor usage.

GENERAL MAINTENANCE

GENERAL MAINTENANCE - GASKET AND O-RINGS

The instrument utilizes gasket and o-ring seals to prevent water from 
entering the battery compartment and the sensor port. Following the 
recommended procedures will help keep the instrument functioning 
properly.  

If the gasket, o-rings and sealing surfaces are not maintained properly, 
it is possible that water can enter the battery compartment and/
or sensor port of the instrument.  If water enters these areas, it can 
damage the battery terminals or sensor port causing loss of battery 
power, false readings and corrosion to the sensors, ports or battery 
terminals. Therefore, when the battery compartment lid is removed, 
the gasket that provides the seal should be carefully inspected for 
contamination (i.e. debris, grit, etc.) and cleaned with water and mild 
detergent if necessary.  

The same inspection should be made of the o-rings associated with 
the pH, ORP and DO sensor connector when replacing a sensor. The 
o-rings should be free of dirt or debris before installing the sensor 
onto the cable. If no dirt or damage to the o-rings is evident, then they 
should be lightly greased with the o-ring grease provided without 
removing them from their groove. However, if there is any indication of 
damage, the sensor o-ring should be replaced with an identical o-ring. 

To remove the sensor o-rings:
Use a small, flat-bladed screwdriver or similar blunt-tipped tool to 
remove the o-ring from its groove near the sensor connector. Check 
the o-ring and the groove for any excess grease or contamination. If 
contamination is evident, clean the o-ring and nearby plastic parts with 
lens cleaning tissue or equivalent lint-free cloth. Alcohol can be used 

to clean the plastic parts, but use only water and mild detergent on the 
o-ring itself. Also, inspect the o-rings for nicks and imperfections. 

Using alcohol on o-rings may cause a loss of elasticity and may 
promote cracking. Do not use a sharp object to remove the 
o-rings.  Damage to the o-ring or the groove may result.

To re-install the sensor o-rings:
Place a small amount of o-ring grease between your thumb and index 
finger.  (More grease is not better!) Draw the o-ring through the grease 
while pressing the fingers together to place a very light covering of 
grease on the o-ring. Place the o-ring into its groove ensuring that it 
does not twist or roll. Use the previously grease-coated finger to once 
again lightly go over the surface of the o-ring.

Do not over-grease the o-rings. Excess grease may collect 
particles that can compromise the seal. Excess grease can also 
cause the waterproofing capabilities of the o-ring to diminish, 
potentially causing leaks. If excess grease is present, remove it 
using a lens cloth or lint-free cloth.

GENERAL MAINTENANCE - SENSOR PORTS

It is important that the entire sensor connector end be dry when 
installing, removing or replacing a sensor. This will prevent water 
from entering the port. Once the sensor is removed, examine the 
connector inside the port. If any moisture is present, use compressed 
air to completely dry the connector or let it air dry. If the connector 
is corroded, contact YSI Technical Support or the YSI authorized 
distributor where you purchased the instrument.  

Remove sensors upside down (facing the ground) to help 
prevent water from entering the port upon removal.

SENSOR MAINTENANCE

Typical working life for pH and ORP sensors is approximately 
12-24 months depending on usage, storage and maintenance. 
Proper storage and maintenance generally extends the sensor’s 
working life. 

i

i

i

i

Содержание YSI Pro1020

Страница 1: ...USER MANUAL English Fran ais Espa ol Portugu s Pro1020 User Manual DOCUMENT 605187...

Страница 2: ...1 Battery Installation 1 Key Pad 2 Connecting the Sensor and Cable 3 Membrane Installation 6 Run Screen 7 Backlight 7 Powering Off 7 Navigation 7 First Power On 8 System Setup Menu 9 DO Local 10 Last...

Страница 3: ...YSI will repair or replace at its sole discretion free of charge any product that YSI determines to be covered by this warranty To exercise this warranty call your local YSI representative or contact...

Страница 4: ...arts or materials are damaged or missing contact YSI Customer Service at 800 897 4151 1 937 767 7241 or the authorized YSI distributor from whom the instrument was purchased BATTERY INSTALLATION The i...

Страница 5: ...he bottom of the Run screen and to increase numerical inputs 3 Power and Backlight Press once to turn instrument on Press a second time to turn backlight on Press a third time to turn backlight off Pr...

Страница 6: ...Next remove the o ring and slide the lid off the sensor INSTALLING THE DISSOLVED OXYGEN SENSOR Before installing a sensor or connecting the cable to the instrument the Sensor Type must be configured f...

Страница 7: ...cap from the dissolved oxygen sensor by pulling it straight off Thoroughly rinse the sensor tip with distilled or deionized water Fill the cap membrane 3 4 full of electrolyte solution then tap the ca...

Страница 8: ...dissolved oxygen sensor type and membrane type Use the up or down arrow keys to highlight the appropriate language sensor and membrane then press enter to confirm figure 6 The Sensor Type must be con...

Страница 9: ...instead of during calibration Example Instrument 1 with DO Local enabled At 737 mmHg barometric pressure roughly 841 ft above sea level the instrument would calibrate to 100 When taking measurements...

Страница 10: ...ompatible sensors for use with a field cable Polarographic This sensor has a black sensor body and is engraved with the model number 2003 Polarographic will be abbreviated Polaro in the instrument men...

Страница 11: ...lay steadily and the instrument will beep twice if Audio is turned on DO Auto Stable can be set to a change of 0 0 to 1 9 over 3 to 19 seconds ISE Auto Stable can be set to a change of 0 0 to 9 9 over...

Страница 12: ...nge the language Highlight the desired language English Spanish Portuguese or French and press enter to enable The enabled language will have an X in the box next to it Highlight the ESC Exit box and...

Страница 13: ...ate formula for this uncorrection is True BP Corrected BP 2 5 Local Altitude in ft above sea level 100 Although the Pro1020 barometer range is 400 0 to 999 9 mmHg you can not adjust the value across t...

Страница 14: ...ubleshooting guide for possible solutions CALIBRATING IN PERCENT DO LOCAL ENABLED Perform this calibration procedure when DO Local is enabled and Quick DO Cal is disabled in the System Setup menu 1 Pe...

Страница 15: ...5 If necessary use the up and down arrow keys to adjust the pH buffer value Note the pH mV reading which ideally should be between 50 and 50 in buffer 7 6 Press enter to continue to second point 7 Ri...

Страница 16: ...membrane If using the 1001A amplified pH sensor be sure to use the guard extender supplied in the 605216 amplified pH sensor kit The extender is screwed onto the sensor bulkhead before the guard or c...

Страница 17: ...enter to regain access to scrolling through the saved data The data set displayed is indicated by the data set number 01 through 50 ERASING DATA While viewing saved data press the enter key to access...

Страница 18: ...or connector Check the o ring and the groove for any excess grease or contamination If contamination is evident clean the o ring and nearby plastic parts with lens cleaning tissue or equivalent lint f...

Страница 19: ...the sensor in clean water wipe with a cotton swab saturated with clean water not DI water and then rerinse with clean water To be certain that all traces of the acid are removed i from the sensor cre...

Страница 20: ...d in the membrane kit Do not sand the electrodes every membrane change as this is not routine maintenance In fact visually the anode may appear tarnished and operate properly YSI recommends using the...

Страница 21: ...embrane should be changed if a bubbles are visible under the membrane b dried electrolyte is visible around the membrane and c if the sensor shows unstable readings or other sensor related symptoms Th...

Страница 22: ...em for service Symptom Possible Solution Instrument will not calibrate pH or ORP instrument displays Calibration Over Calibration Under or Unstable Reading during calibration 1 Verify correct ISE sens...

Страница 23: ...specifications represent typical performance and are subject to change without notice For the latest product specification information please visit YSI s website at ysi com or contact YSI Tech Suppor...

Страница 24: ...E blue cap 6 each includes electrolyte Used with Galvanic sensor All cables include a temperature sensor The pH or ORP and Dissolved Oxygen sensors are sold separately D E C L A R AT I O N O F C O N F...

Страница 25: ...country and region and users are expected to understand and follow the battery disposal requirements for their specific locale C O N TA C T I N F O R M AT I O N ORDERING AND TECHNICAL SUPPORT Telephon...

Страница 26: ...Item 605187 Rev A January 2014 2014 YSI Incorporated...

Отзывы: