36

37

bottle that was originally included with the sensor. This can be filled 
with buffer 4 solution. If you no longer have the storage bottle, simply 
place the sensor in a buffer 4 solution.

When storing the dissolved oxygen sensor for more than 30 days, 
remove the cap membrane and thoroughly rinse the dissolved 
oxygen sensor with distilled or deionized water. Once the sensor 
has been rinsed either blow it dry with compressed air or allow to air 
dry completely. Next, install a new, clean, dry cap membrane onto 
the sensor. This will keep the sensor dry and protect the anode and 
cathode during storage. After storing for a long period of time, it is 
necessary to “condition” the dissolved oxygen sensor by installing a 
new membrane with electrolyte solution.

Long Term Storage Temp: -5 to 70°C (23 to 158°F) without pH
 

 

 

      0 to 30°C (32 to 86°F) with pH or ORP*

*Operating temperature range for pH sensor is -5 to 60°C (23 to 140°F) 
and operating temperature range for the ORP sensor is 0 to 60°C (32 
to 140°F) 

T R O U B L E S H O O T I N G

Symptom

Possible Solution

Instrument will 
not turn on, a 
battery symbol  
appears, or “Critical 
Shutdown” displays 
on the screen.

1.  Low battery voltage, replace batteries.
2.  Batteries installed incorrectly, 

check battery polarity.

3.  Return system for service.

Temperature values 
display Over or Undr 
on Run screen.

1.  Sample temperature is less than -5° C or more 

than +55°C. Increase or decrease the sample 
temperature to bring within the allowable 
range.

2.  Contact YSI Tech Support. Possible temperature 

sensor failure.

Barometer reads 
over/undr and 
calibrating DO results 
in a Barometric 
Pressure Over/Undr 
error message.

1.  Barometer failure, return system for service

Symptom

Possible Solution

Instrument will not 
calibrate pH or ORP; 
instrument displays 
“Calibration Over”, 
“Calibration Under”, 
or “Unstable Reading” 
during calibration.

1.  Verify correct ISE sensor type selection 

in the System Setup menu.  

2.  Verify the calibration solution is accurate.
3.  If calibrating pH, make sure you 

are calibrating buffer 7 first. 

4.  Clean the pH or ORP sensor.
5.  Contact YSI Tech Support.

pH or ORP readings 
are inaccurate.

1.  Verify correct ISE sensor type selection 

in the System Setup menu.

2.  Verify temperature readings are accurate.
3.  Recalibrate the pH or ORP sensor.
4.  Clean the pH or ORP sensor.
5.  Contact YSI Tech Support.

pH values display 
Over or Undr on 
Run screen. 

1.  Verify correct ISE sensor type selection 

in the System Setup menu. 

2.  Verify temperature readings are accurate.
3.  Recalibrate the pH sensor. 
4.  Clean the pH sensor and recalibrate.
5.  Contact YSI Tech Support.

ORP values display 
Over or Undr on 
Run screen. 

1.  Verify correct sensor type selection 

in the System Setup menu.

2.  Sample ORP value is outside the 

measurement range of -1500 to 1500 mV. 

3.  Verify temperature readings are accurate.
4.  Recalibrate the ORP sensor. 
5.  Clean the ORP sensor and recalibrate.
6.  Contact YSI Tech Support.

Instrument will not 
calibrate dissolved 
oxygen; instrument 
displays “Calibration 
Over”, “Calibration 
Under”, or “Unstable 
Reading” during 
calibration.

1.  Verify barometer reading.
2.  Verify correct DO sensor and DO membrane 

type selection in the System Setup menu.  

3.  Calibrations cup may not be 100% 

water saturated, see DO Calibration.

4.  Ensure adequate sample movement if 

performing mg/L or ppm calibration.

5.  Allow sufficient stabilization time for 

dissolved oxygen and temperature AND 
wait at least 3 seconds before confirming 
a DO% or DO Local% calibration.

6.  Replace membrane and electrolyte.
7.  Clean sensor electrodes.
8.  Contact YSI Tech Support.

Содержание YSI Pro1020

Страница 1: ...USER MANUAL English Fran ais Espa ol Portugu s Pro1020 User Manual DOCUMENT 605187...

Страница 2: ...1 Battery Installation 1 Key Pad 2 Connecting the Sensor and Cable 3 Membrane Installation 6 Run Screen 7 Backlight 7 Powering Off 7 Navigation 7 First Power On 8 System Setup Menu 9 DO Local 10 Last...

Страница 3: ...YSI will repair or replace at its sole discretion free of charge any product that YSI determines to be covered by this warranty To exercise this warranty call your local YSI representative or contact...

Страница 4: ...arts or materials are damaged or missing contact YSI Customer Service at 800 897 4151 1 937 767 7241 or the authorized YSI distributor from whom the instrument was purchased BATTERY INSTALLATION The i...

Страница 5: ...he bottom of the Run screen and to increase numerical inputs 3 Power and Backlight Press once to turn instrument on Press a second time to turn backlight on Press a third time to turn backlight off Pr...

Страница 6: ...Next remove the o ring and slide the lid off the sensor INSTALLING THE DISSOLVED OXYGEN SENSOR Before installing a sensor or connecting the cable to the instrument the Sensor Type must be configured f...

Страница 7: ...cap from the dissolved oxygen sensor by pulling it straight off Thoroughly rinse the sensor tip with distilled or deionized water Fill the cap membrane 3 4 full of electrolyte solution then tap the ca...

Страница 8: ...dissolved oxygen sensor type and membrane type Use the up or down arrow keys to highlight the appropriate language sensor and membrane then press enter to confirm figure 6 The Sensor Type must be con...

Страница 9: ...instead of during calibration Example Instrument 1 with DO Local enabled At 737 mmHg barometric pressure roughly 841 ft above sea level the instrument would calibrate to 100 When taking measurements...

Страница 10: ...ompatible sensors for use with a field cable Polarographic This sensor has a black sensor body and is engraved with the model number 2003 Polarographic will be abbreviated Polaro in the instrument men...

Страница 11: ...lay steadily and the instrument will beep twice if Audio is turned on DO Auto Stable can be set to a change of 0 0 to 1 9 over 3 to 19 seconds ISE Auto Stable can be set to a change of 0 0 to 9 9 over...

Страница 12: ...nge the language Highlight the desired language English Spanish Portuguese or French and press enter to enable The enabled language will have an X in the box next to it Highlight the ESC Exit box and...

Страница 13: ...ate formula for this uncorrection is True BP Corrected BP 2 5 Local Altitude in ft above sea level 100 Although the Pro1020 barometer range is 400 0 to 999 9 mmHg you can not adjust the value across t...

Страница 14: ...ubleshooting guide for possible solutions CALIBRATING IN PERCENT DO LOCAL ENABLED Perform this calibration procedure when DO Local is enabled and Quick DO Cal is disabled in the System Setup menu 1 Pe...

Страница 15: ...5 If necessary use the up and down arrow keys to adjust the pH buffer value Note the pH mV reading which ideally should be between 50 and 50 in buffer 7 6 Press enter to continue to second point 7 Ri...

Страница 16: ...membrane If using the 1001A amplified pH sensor be sure to use the guard extender supplied in the 605216 amplified pH sensor kit The extender is screwed onto the sensor bulkhead before the guard or c...

Страница 17: ...enter to regain access to scrolling through the saved data The data set displayed is indicated by the data set number 01 through 50 ERASING DATA While viewing saved data press the enter key to access...

Страница 18: ...or connector Check the o ring and the groove for any excess grease or contamination If contamination is evident clean the o ring and nearby plastic parts with lens cleaning tissue or equivalent lint f...

Страница 19: ...the sensor in clean water wipe with a cotton swab saturated with clean water not DI water and then rerinse with clean water To be certain that all traces of the acid are removed i from the sensor cre...

Страница 20: ...d in the membrane kit Do not sand the electrodes every membrane change as this is not routine maintenance In fact visually the anode may appear tarnished and operate properly YSI recommends using the...

Страница 21: ...embrane should be changed if a bubbles are visible under the membrane b dried electrolyte is visible around the membrane and c if the sensor shows unstable readings or other sensor related symptoms Th...

Страница 22: ...em for service Symptom Possible Solution Instrument will not calibrate pH or ORP instrument displays Calibration Over Calibration Under or Unstable Reading during calibration 1 Verify correct ISE sens...

Страница 23: ...specifications represent typical performance and are subject to change without notice For the latest product specification information please visit YSI s website at ysi com or contact YSI Tech Suppor...

Страница 24: ...E blue cap 6 each includes electrolyte Used with Galvanic sensor All cables include a temperature sensor The pH or ORP and Dissolved Oxygen sensors are sold separately D E C L A R AT I O N O F C O N F...

Страница 25: ...country and region and users are expected to understand and follow the battery disposal requirements for their specific locale C O N TA C T I N F O R M AT I O N ORDERING AND TECHNICAL SUPPORT Telephon...

Страница 26: ...Item 605187 Rev A January 2014 2014 YSI Incorporated...

Отзывы: