background image

20

21

If using a Galvanic sensor, wait approximately 5 to 10 minutes 
for the cup to become completely saturated. The instrument’s 
Auto Shutoff should be disabled or set to at least 20 minutes. 

4.  Ensure the barometer is reading accurately. If necessary, 

perform a barometer calibration.

5.  Press and hold the Calibrate key for 3 seconds. Highlight 

Dissolved Oxygen and press enter. The Pro1020 will indicate 
‘Calibrating %DO’ on the display. The instrument will 
automatically calibrate the sensor to the current barometric 
pressure. If DO Local% is enabled, the sensor will calibrate to 
100%. This may take up to 2 minutes depending on the age 
and condition of the sensor and  membrane.  You can press the 
Cal key at this time to cancel the calibration if needed. 

6.  ‘Calibration Successful’ will display for a few seconds to indicate 

a successful calibration and then the instrument will return to 
the Run screen.

7.  If the calibration is unsuccessful, an error message will display 

on the screen. Press the Cal key to exit the error message and 
return to the Run screen. See the Troubleshooting guide for 
possible solutions.  

CALIBRATING IN PERCENT (DO%)   

 

Perform this calibration procedure when Quick DO Cal is disabled in 
the System Setup menu.

1.  Perform steps 1-4 of the Quick DO Calibration procedure.
2.  Press and hold the Calibrate key for 3 seconds. Highlight 

Dissolved Oxygen and press enter. Next, highlight % and press 
enter.  

3.  The Pro1020 will display the current DO% and temperature 

readings along with the % calibration value.  The % calibration 
value is based on the barometer reading. 

4.  Wait at least 3 seconds, then, once the DO% and temperature 

readings are stable, press enter to complete the calibration.  
Or, press the Cal key to cancel the calibration.    

5.  ‘Calibration Successful’ will display for a few seconds to indicate 

a successful calibration and then the instrument will return to 
the Run screen.

6.  If the calibration is unsuccessful, an error message will display 

on the screen. Press the Cal key to exit the calibration error 
message and return to the Run screen.  See the Troubleshooting 
guide for possible solutions.  

CALIBRATING IN PERCENT (DO LOCAL% ENABLED)  

Perform this calibration procedure when DO Local% is enabled and 
Quick DO Cal is disabled in the System Setup menu.  

1.  Perform steps 1-4 of the Quick DO Calibration procedure.
2.  Press and hold the Cal key for 3 seconds. Highlight Dissolved 

Oxygen and press enter. 

3.  %Local will automatically be highlighted, press enter. The 

Pro1020 will display the current DO% and temperature 
readings along with the % calibration value. The % calibration 
value will always be 100% for DO Local%.  

4.  Wait at least 3 seconds, then, once the DO%L and temperature 

readings stabilize, press enter to complete the calibration. Or, 
press the Cal key to cancel the calibration.  

5.  ‘Calibration Successful’ will display for a few seconds to indicate 

a successful calibration and then the instrument will return to 
the Run screen.

6.  If the calibration is unsuccessful, an error message will display 

on the screen. Press the Cal key to exit the calibration error 
message and return to the Run screen.  See the Troubleshooting 
guide for possible solutions.  

CALIBRATING IN MG/L OR PPM    

1.  Power the instrument on and place the sensor into a sample 

that has been titrated to determine the dissolved oxygen 
concentration. The dissolved oxygen and temperature sensor 
should be in the sample. During the calibration, continuously 
stir or move the probe in the sample at a rate of at least 6 inches 
(15.5 cm) per second if using a yellow membrane or at least 3 
inches (7.7 cm) per second if using a blue membrane.  A stir 
plate may be helpful for this calibration procedure.  

2.  Allow the dissolved oxygen and temperature readings to 

stabilize. This may take 5 to 15 minutes, depending on the type 
and condition of the sensor.    

3.  Press the Cal key.  Highlight Dissolved Oxygen and press enter. 
4.  Highlight mg/L or ppm depending on what is enabled in the 

System Setup menu and press enter.  

5.  Use the up and down arrow keys to adjust the mg/L (ppm) 

reading to the value of the titrated sample. Press enter to 
confirm the value and complete the calibration or press the 
Cal key to cancel the calibration.

6.  ‘Calibration Successful’ will display for a few seconds to indicate 

a successful calibration and then the instrument will return to 
the Run screen.

Содержание YSI Pro1020

Страница 1: ...USER MANUAL English Fran ais Espa ol Portugu s Pro1020 User Manual DOCUMENT 605187...

Страница 2: ...1 Battery Installation 1 Key Pad 2 Connecting the Sensor and Cable 3 Membrane Installation 6 Run Screen 7 Backlight 7 Powering Off 7 Navigation 7 First Power On 8 System Setup Menu 9 DO Local 10 Last...

Страница 3: ...YSI will repair or replace at its sole discretion free of charge any product that YSI determines to be covered by this warranty To exercise this warranty call your local YSI representative or contact...

Страница 4: ...arts or materials are damaged or missing contact YSI Customer Service at 800 897 4151 1 937 767 7241 or the authorized YSI distributor from whom the instrument was purchased BATTERY INSTALLATION The i...

Страница 5: ...he bottom of the Run screen and to increase numerical inputs 3 Power and Backlight Press once to turn instrument on Press a second time to turn backlight on Press a third time to turn backlight off Pr...

Страница 6: ...Next remove the o ring and slide the lid off the sensor INSTALLING THE DISSOLVED OXYGEN SENSOR Before installing a sensor or connecting the cable to the instrument the Sensor Type must be configured f...

Страница 7: ...cap from the dissolved oxygen sensor by pulling it straight off Thoroughly rinse the sensor tip with distilled or deionized water Fill the cap membrane 3 4 full of electrolyte solution then tap the ca...

Страница 8: ...dissolved oxygen sensor type and membrane type Use the up or down arrow keys to highlight the appropriate language sensor and membrane then press enter to confirm figure 6 The Sensor Type must be con...

Страница 9: ...instead of during calibration Example Instrument 1 with DO Local enabled At 737 mmHg barometric pressure roughly 841 ft above sea level the instrument would calibrate to 100 When taking measurements...

Страница 10: ...ompatible sensors for use with a field cable Polarographic This sensor has a black sensor body and is engraved with the model number 2003 Polarographic will be abbreviated Polaro in the instrument men...

Страница 11: ...lay steadily and the instrument will beep twice if Audio is turned on DO Auto Stable can be set to a change of 0 0 to 1 9 over 3 to 19 seconds ISE Auto Stable can be set to a change of 0 0 to 9 9 over...

Страница 12: ...nge the language Highlight the desired language English Spanish Portuguese or French and press enter to enable The enabled language will have an X in the box next to it Highlight the ESC Exit box and...

Страница 13: ...ate formula for this uncorrection is True BP Corrected BP 2 5 Local Altitude in ft above sea level 100 Although the Pro1020 barometer range is 400 0 to 999 9 mmHg you can not adjust the value across t...

Страница 14: ...ubleshooting guide for possible solutions CALIBRATING IN PERCENT DO LOCAL ENABLED Perform this calibration procedure when DO Local is enabled and Quick DO Cal is disabled in the System Setup menu 1 Pe...

Страница 15: ...5 If necessary use the up and down arrow keys to adjust the pH buffer value Note the pH mV reading which ideally should be between 50 and 50 in buffer 7 6 Press enter to continue to second point 7 Ri...

Страница 16: ...membrane If using the 1001A amplified pH sensor be sure to use the guard extender supplied in the 605216 amplified pH sensor kit The extender is screwed onto the sensor bulkhead before the guard or c...

Страница 17: ...enter to regain access to scrolling through the saved data The data set displayed is indicated by the data set number 01 through 50 ERASING DATA While viewing saved data press the enter key to access...

Страница 18: ...or connector Check the o ring and the groove for any excess grease or contamination If contamination is evident clean the o ring and nearby plastic parts with lens cleaning tissue or equivalent lint f...

Страница 19: ...the sensor in clean water wipe with a cotton swab saturated with clean water not DI water and then rerinse with clean water To be certain that all traces of the acid are removed i from the sensor cre...

Страница 20: ...d in the membrane kit Do not sand the electrodes every membrane change as this is not routine maintenance In fact visually the anode may appear tarnished and operate properly YSI recommends using the...

Страница 21: ...embrane should be changed if a bubbles are visible under the membrane b dried electrolyte is visible around the membrane and c if the sensor shows unstable readings or other sensor related symptoms Th...

Страница 22: ...em for service Symptom Possible Solution Instrument will not calibrate pH or ORP instrument displays Calibration Over Calibration Under or Unstable Reading during calibration 1 Verify correct ISE sens...

Страница 23: ...specifications represent typical performance and are subject to change without notice For the latest product specification information please visit YSI s website at ysi com or contact YSI Tech Suppor...

Страница 24: ...E blue cap 6 each includes electrolyte Used with Galvanic sensor All cables include a temperature sensor The pH or ORP and Dissolved Oxygen sensors are sold separately D E C L A R AT I O N O F C O N F...

Страница 25: ...country and region and users are expected to understand and follow the battery disposal requirements for their specific locale C O N TA C T I N F O R M AT I O N ORDERING AND TECHNICAL SUPPORT Telephon...

Страница 26: ...Item 605187 Rev A January 2014 2014 YSI Incorporated...

Отзывы: