background image

12

13

When Quick DO Cal is enabled, press and hold the calibration 

 

key for 3 seconds while in the Run screen. Next, highlight Dissolved 
Oxygen and press enter to calibrate the DO sensor to the instrument’s 
barometer reading. For more information on Quick DO Cal, see the 
Calibration section of this manual. 

AUDIO  

Audio can be enabled by highlighting Audio and pressing enter. When 
enabled, there will be an ‘X’ in the box next to Audio. 

When Audio is enabled, the Pro1020 will beep twice to indicate stability 
when Auto Stable is enabled. The instrument will also beep when a key 
is pressed.  When Audio is disabled, the Pro1020 will not beep. 

CONTRAST

To adjust the display Contrast, use the up or down arrow keys to 
highlight Contrast, then press enter. Next, use the up or down arrow 
keys to adjust the contrast.  The up arrow key will darken the contrast 
and the down arrow key will lighten the contrast. After adjusting the 
contrast, press enter to save and exit the Contrast adjustment function.   

ALTERNATE CONTRAST ADJUSTMENT OPTION

If necessary, there is an alternate method of adjusting the contrast. To 
adjust the contrast, press and hold the menu key, then press the up 
arrow key to darken the contrast or press the down arrow key to lighten 
the contrast.  

DO SENSOR TYPE

DO Sensor Type sets the type of dissolved oxygen sensor being used; 
either polarographic (black) or galvanic (grey).

Use the up or down arrow keys to highlight Sensor Type, then press 
enter to open a submenu. Highlight the sensor type corresponding 
to the sensor installed on the cable and press enter to confirm. The 
enabled sensor type will have an ‘X’ in the box next to it. Next, use the 
down arrow key to highlight the ESC – Exit box, then press enter to 
save changes and to close the sensor submenu.  

The Pro1020 has two compatible sensors for use with a field cable: 

• 

Polarographic – This sensor has a black sensor body and is engraved 
with the model number 2003. Polarographic will be abbreviated 
Polaro in the instrument menu.

• 

Galvanic – This sensor has a grey sensor body and is engraved with 
the model number 2002. 

In terms of physical configuration, membrane material and general 
performance, YSI Pro Series galvanic sensors are exactly like the Pro 
Series polarographic sensors. The advantage of using galvanic sensors 
is convenience. Galvanic sensors provide an instant-on sensor without 
the need for warm-up time but this affects the life of the sensor. 
Polarographic sensors last longer and have a longer warranty but 
require a 5-15 minute warm-up time before use or calibration.  

MEMBRANE TYPE

Membrane Type sets the type of membrane used on the dissolved 
oxygen sensor; either 1.25 PE (Yellow) or 2.0 PE (blue). Use the up 
or down arrow keys to highlight Membrane Type and press enter to 
open the submenu. Highlight the membrane type corresponding to 
the membrane installed on the sensor and press enter to confirm. The 
enabled membrane type will have an ‘X’ in the box next to it. Use the 
down arrow key to highlight the ESC – Exit box and press enter to save 
changes and to close the membrane submenu.    

The dissolved oxygen sensor is supplied with membranes specific 
to the sensor type and are color coded as described in the following 
tables.  

Galvanic Membrane Kits: 

Model

Color

Material 

Description

5913

Yellow 1.25 mil 

polyethylene (PE)

Faster response time and less flow 
dependence than traditional FEP 
membranes.  

5914

Blue

2.0 mil 
polyethylene (PE)

Less flow dependence than 1.25 mil 
membrane but little slower response. 

Polarographic Membrane Kits: 

Item

Color

Material 

Description

5908 Yellow 1.25 mil 

polyethylene (PE)

Faster response time and less flow 
dependence than traditional FEP 
membranes.  

5909

Blue

2.0 mil 
polyethylene (PE)

Less flow dependence than 1.25 mil 
membrane but a little slower response. 

Содержание YSI Pro1020

Страница 1: ...USER MANUAL English Fran ais Espa ol Portugu s Pro1020 User Manual DOCUMENT 605187...

Страница 2: ...1 Battery Installation 1 Key Pad 2 Connecting the Sensor and Cable 3 Membrane Installation 6 Run Screen 7 Backlight 7 Powering Off 7 Navigation 7 First Power On 8 System Setup Menu 9 DO Local 10 Last...

Страница 3: ...YSI will repair or replace at its sole discretion free of charge any product that YSI determines to be covered by this warranty To exercise this warranty call your local YSI representative or contact...

Страница 4: ...arts or materials are damaged or missing contact YSI Customer Service at 800 897 4151 1 937 767 7241 or the authorized YSI distributor from whom the instrument was purchased BATTERY INSTALLATION The i...

Страница 5: ...he bottom of the Run screen and to increase numerical inputs 3 Power and Backlight Press once to turn instrument on Press a second time to turn backlight on Press a third time to turn backlight off Pr...

Страница 6: ...Next remove the o ring and slide the lid off the sensor INSTALLING THE DISSOLVED OXYGEN SENSOR Before installing a sensor or connecting the cable to the instrument the Sensor Type must be configured f...

Страница 7: ...cap from the dissolved oxygen sensor by pulling it straight off Thoroughly rinse the sensor tip with distilled or deionized water Fill the cap membrane 3 4 full of electrolyte solution then tap the ca...

Страница 8: ...dissolved oxygen sensor type and membrane type Use the up or down arrow keys to highlight the appropriate language sensor and membrane then press enter to confirm figure 6 The Sensor Type must be con...

Страница 9: ...instead of during calibration Example Instrument 1 with DO Local enabled At 737 mmHg barometric pressure roughly 841 ft above sea level the instrument would calibrate to 100 When taking measurements...

Страница 10: ...ompatible sensors for use with a field cable Polarographic This sensor has a black sensor body and is engraved with the model number 2003 Polarographic will be abbreviated Polaro in the instrument men...

Страница 11: ...lay steadily and the instrument will beep twice if Audio is turned on DO Auto Stable can be set to a change of 0 0 to 1 9 over 3 to 19 seconds ISE Auto Stable can be set to a change of 0 0 to 9 9 over...

Страница 12: ...nge the language Highlight the desired language English Spanish Portuguese or French and press enter to enable The enabled language will have an X in the box next to it Highlight the ESC Exit box and...

Страница 13: ...ate formula for this uncorrection is True BP Corrected BP 2 5 Local Altitude in ft above sea level 100 Although the Pro1020 barometer range is 400 0 to 999 9 mmHg you can not adjust the value across t...

Страница 14: ...ubleshooting guide for possible solutions CALIBRATING IN PERCENT DO LOCAL ENABLED Perform this calibration procedure when DO Local is enabled and Quick DO Cal is disabled in the System Setup menu 1 Pe...

Страница 15: ...5 If necessary use the up and down arrow keys to adjust the pH buffer value Note the pH mV reading which ideally should be between 50 and 50 in buffer 7 6 Press enter to continue to second point 7 Ri...

Страница 16: ...membrane If using the 1001A amplified pH sensor be sure to use the guard extender supplied in the 605216 amplified pH sensor kit The extender is screwed onto the sensor bulkhead before the guard or c...

Страница 17: ...enter to regain access to scrolling through the saved data The data set displayed is indicated by the data set number 01 through 50 ERASING DATA While viewing saved data press the enter key to access...

Страница 18: ...or connector Check the o ring and the groove for any excess grease or contamination If contamination is evident clean the o ring and nearby plastic parts with lens cleaning tissue or equivalent lint f...

Страница 19: ...the sensor in clean water wipe with a cotton swab saturated with clean water not DI water and then rerinse with clean water To be certain that all traces of the acid are removed i from the sensor cre...

Страница 20: ...d in the membrane kit Do not sand the electrodes every membrane change as this is not routine maintenance In fact visually the anode may appear tarnished and operate properly YSI recommends using the...

Страница 21: ...embrane should be changed if a bubbles are visible under the membrane b dried electrolyte is visible around the membrane and c if the sensor shows unstable readings or other sensor related symptoms Th...

Страница 22: ...em for service Symptom Possible Solution Instrument will not calibrate pH or ORP instrument displays Calibration Over Calibration Under or Unstable Reading during calibration 1 Verify correct ISE sens...

Страница 23: ...specifications represent typical performance and are subject to change without notice For the latest product specification information please visit YSI s website at ysi com or contact YSI Tech Suppor...

Страница 24: ...E blue cap 6 each includes electrolyte Used with Galvanic sensor All cables include a temperature sensor The pH or ORP and Dissolved Oxygen sensors are sold separately D E C L A R AT I O N O F C O N F...

Страница 25: ...country and region and users are expected to understand and follow the battery disposal requirements for their specific locale C O N TA C T I N F O R M AT I O N ORDERING AND TECHNICAL SUPPORT Telephon...

Страница 26: ...Item 605187 Rev A January 2014 2014 YSI Incorporated...

Отзывы: