background image

8

9

   NAVIGATING THE RUN SCREEN

When in the Run screen, the up   and down  arrow keys will move the 
highlighted box along the bottom options. Once a box is highlighted, 
press enter to access the highlighted option. 

Description of Run screen box functions from left to right:

Option

Description

Barometer

Highlight and press enter to calibrate the 
instrument’s internal barometer.

Salinity Value

Highlight and press enter to adjust the salinity 
compensation value that is used for the 
dissolved oxygen mg/L measurement. 

SAVE

Highlight and press enter to save displayed 
data to memory.

DATA

Highlight and press enter to view and/or erase 
saved data.

NAVIGATING THE SYSTEM SETUP MENU

When in the System Setup menu, the up and down arrow keys will 
move the highlighted bar up and down the system setup options. See 
the System Setup menu section of this manual for more information 
about these options.  

FIRST POWER ON

The instrument will step through an initial configuration when powered 
on for the first time. This will set the language, dissolved oxygen sensor 
type and membrane type. Use the up or down arrow keys to highlight 
the appropriate language, sensor and membrane, then press enter 
to confirm (figure 6). The Sensor Type must be configured for the 
dissolved oxygen sensor installed on the cable. Failure to do this may 
result in damage not covered under warranty.  If an incorrect option is 
selected, it may be changed in the System Setup menu.

 

 

Select Sensor Type: 

 

 

Polaro (black) 

 

Galvanic (grey) 

 

 

 

Use 



 to select 

sensor type 

Press 

  to confirm 

 

 

 

Select Membrane 

Type: 

 

   

1.25 (Yellow) 

   

 2.0 (Blue)

 

 

 

Use 



 to select 

membrane 

Press 

  to confirm 

 

  

Figure 6, Select language, dissolved oxygen sensor and membrane.

After selecting a language, sensor and membrane, the Run screen will 
be displayed. The next time the instrument is powered up, the Run 
screen will display immediately after the splash screen. If the sensor 
type or membrane type is changed, ensure that it is updated in the 
System Setup menu. 

S Y S T E M   S E T U P   M E N U

Press the menu 

  key to access the System Setup menu. The System 

Setup menu contains three screens notated as ‘pages’.  The current 
page is indicated near the bottom of the display (figure 7).  Use the up 
and down arrow keys to scroll through menu options and menu pages.  

Содержание YSI Pro1020

Страница 1: ...USER MANUAL English Fran ais Espa ol Portugu s Pro1020 User Manual DOCUMENT 605187...

Страница 2: ...1 Battery Installation 1 Key Pad 2 Connecting the Sensor and Cable 3 Membrane Installation 6 Run Screen 7 Backlight 7 Powering Off 7 Navigation 7 First Power On 8 System Setup Menu 9 DO Local 10 Last...

Страница 3: ...YSI will repair or replace at its sole discretion free of charge any product that YSI determines to be covered by this warranty To exercise this warranty call your local YSI representative or contact...

Страница 4: ...arts or materials are damaged or missing contact YSI Customer Service at 800 897 4151 1 937 767 7241 or the authorized YSI distributor from whom the instrument was purchased BATTERY INSTALLATION The i...

Страница 5: ...he bottom of the Run screen and to increase numerical inputs 3 Power and Backlight Press once to turn instrument on Press a second time to turn backlight on Press a third time to turn backlight off Pr...

Страница 6: ...Next remove the o ring and slide the lid off the sensor INSTALLING THE DISSOLVED OXYGEN SENSOR Before installing a sensor or connecting the cable to the instrument the Sensor Type must be configured f...

Страница 7: ...cap from the dissolved oxygen sensor by pulling it straight off Thoroughly rinse the sensor tip with distilled or deionized water Fill the cap membrane 3 4 full of electrolyte solution then tap the ca...

Страница 8: ...dissolved oxygen sensor type and membrane type Use the up or down arrow keys to highlight the appropriate language sensor and membrane then press enter to confirm figure 6 The Sensor Type must be con...

Страница 9: ...instead of during calibration Example Instrument 1 with DO Local enabled At 737 mmHg barometric pressure roughly 841 ft above sea level the instrument would calibrate to 100 When taking measurements...

Страница 10: ...ompatible sensors for use with a field cable Polarographic This sensor has a black sensor body and is engraved with the model number 2003 Polarographic will be abbreviated Polaro in the instrument men...

Страница 11: ...lay steadily and the instrument will beep twice if Audio is turned on DO Auto Stable can be set to a change of 0 0 to 1 9 over 3 to 19 seconds ISE Auto Stable can be set to a change of 0 0 to 9 9 over...

Страница 12: ...nge the language Highlight the desired language English Spanish Portuguese or French and press enter to enable The enabled language will have an X in the box next to it Highlight the ESC Exit box and...

Страница 13: ...ate formula for this uncorrection is True BP Corrected BP 2 5 Local Altitude in ft above sea level 100 Although the Pro1020 barometer range is 400 0 to 999 9 mmHg you can not adjust the value across t...

Страница 14: ...ubleshooting guide for possible solutions CALIBRATING IN PERCENT DO LOCAL ENABLED Perform this calibration procedure when DO Local is enabled and Quick DO Cal is disabled in the System Setup menu 1 Pe...

Страница 15: ...5 If necessary use the up and down arrow keys to adjust the pH buffer value Note the pH mV reading which ideally should be between 50 and 50 in buffer 7 6 Press enter to continue to second point 7 Ri...

Страница 16: ...membrane If using the 1001A amplified pH sensor be sure to use the guard extender supplied in the 605216 amplified pH sensor kit The extender is screwed onto the sensor bulkhead before the guard or c...

Страница 17: ...enter to regain access to scrolling through the saved data The data set displayed is indicated by the data set number 01 through 50 ERASING DATA While viewing saved data press the enter key to access...

Страница 18: ...or connector Check the o ring and the groove for any excess grease or contamination If contamination is evident clean the o ring and nearby plastic parts with lens cleaning tissue or equivalent lint f...

Страница 19: ...the sensor in clean water wipe with a cotton swab saturated with clean water not DI water and then rerinse with clean water To be certain that all traces of the acid are removed i from the sensor cre...

Страница 20: ...d in the membrane kit Do not sand the electrodes every membrane change as this is not routine maintenance In fact visually the anode may appear tarnished and operate properly YSI recommends using the...

Страница 21: ...embrane should be changed if a bubbles are visible under the membrane b dried electrolyte is visible around the membrane and c if the sensor shows unstable readings or other sensor related symptoms Th...

Страница 22: ...em for service Symptom Possible Solution Instrument will not calibrate pH or ORP instrument displays Calibration Over Calibration Under or Unstable Reading during calibration 1 Verify correct ISE sens...

Страница 23: ...specifications represent typical performance and are subject to change without notice For the latest product specification information please visit YSI s website at ysi com or contact YSI Tech Suppor...

Страница 24: ...E blue cap 6 each includes electrolyte Used with Galvanic sensor All cables include a temperature sensor The pH or ORP and Dissolved Oxygen sensors are sold separately D E C L A R AT I O N O F C O N F...

Страница 25: ...country and region and users are expected to understand and follow the battery disposal requirements for their specific locale C O N TA C T I N F O R M AT I O N ORDERING AND TECHNICAL SUPPORT Telephon...

Страница 26: ...Item 605187 Rev A January 2014 2014 YSI Incorporated...

Отзывы: