background image

30

31

SENSOR MAINTENANCE - TEMPERATURE

You must keep the temperature sensor free of build up. No additional 
maintenance is required. A toothbrush can be used to scrub the 
temperature sensor if needed.
 

SENSOR MAINTENANCE - pH AND ORP

CAUTION

: When using a cotton swab, be careful NOT to wedge 

the swab between the guard and the glass sensor. If necessary, 
remove cotton from the swab tip, so that the cotton can reach 
all parts of the sensor tip without stress.  You can also use a pipe 
cleaner for this operation if more convenient. 

Cleaning is required whenever deposits or contaminants appear on 
the glass or platinum sensor surfaces or when the sensor’s response 
slows.  The cleaning can be chemical and/or mechanical.

Removing the sensor from the cable may make cleaning easier.   Initially, 
use clean water and a soft clean cloth, lens cleaning tissue, or cotton 
swab to remove all foreign material from the glass bulb or platinum 
button. Then use a moistened cotton swab to carefully remove any 
material that may be blocking the reference electrode junction of the 
sensor.

If good pH or ORP response is not restored, perform the following 
additional procedure:

1.  Soak the sensor for 10-15 minutes in clean water containing a 

few drops of commercial dish washing liquid.

2.  GENTLY clean the glass bulb and platinum button by rubbing 

with a cotton swab soaked in the cleaning solution.

3.  Rinse the sensor in clean water (not DI water), wipe with a cotton 

swab saturated with clean water, and then rerinse with clean 
water.

If good pH or ORP response is still not restored, perform the following 
additional procedure:

1.  Soak the sensor for 30-60 minutes in one molar (1 M) hydrochloric 

acid (HCl). This reagent can be purchased from most lab supply 
distributors. Be sure to follow the safety instructions included 
with the acid.

2.  Rinse the sensor in clean water, wipe with a cotton swab 

saturated with clean water (not DI water), and then rerinse with 
clean water.  To be certain that all traces of the acid are removed 

i

from the sensor crevices, soak the sensor in clean water for 
about an hour with occasional stirring.

CAUTION

:  Do NOT mix the acid from the previous step with the 

chlorine bleach in the following step. Toxic gaseous products 
can be formed from the reaction between acid and chlorine 
bleach. Be certain to copiously rinse the sink and drain system 
of acid after its disposal and before disposal of the chlorine 
bleach.  

If biological contamination of the reference junction is suspected or if 
good response is not restored by the above procedures, perform the 
following additional cleaning step:

1.  First, be sure all HCl acid from the previous cleaning step has 

been rinsed from the probe, sink and drain! 

2.  Soak the sensor for approximately 1 hour in a 1:1 dilution of 

commercially-available chlorine bleach. 

3.  Rinse the sensor with clean water (not DI water) and then soak 

for at least 1 hour in clean water with occasional stirring to 
remove residual bleach from the junction.  (If possible, soak the 
sensor for a period of time longer than 1 hour in order to be 
certain that all traces of chlorine bleach are removed.)  Then 
rerinse the sensor with clean water and retest.

Dry the port and sensor connector with compressed air and 
apply a very thin coat of o-ring lubricant to all o-rings before 
reinstallation.

If this procedure is unsuccessful, as indicated by improper sensor 
performance, contact YSI Technical Support or the YSI authorized 
dealer where you purchased the instrument.

SENSOR MAINTENANCE - DISSOLVED OXYGEN  

Membrane Cap Installation

The DO sensor (Polarographic or Galvanic) is shipped with a protective 
red cap that needs to be removed before using. Remove the red 
protective cap or used cap membrane and replace it with a new cap 
membrane following these instructions: 

Remove the sensor guard to access the sensor.

i

Содержание YSI Pro1020

Страница 1: ...USER MANUAL English Fran ais Espa ol Portugu s Pro1020 User Manual DOCUMENT 605187...

Страница 2: ...1 Battery Installation 1 Key Pad 2 Connecting the Sensor and Cable 3 Membrane Installation 6 Run Screen 7 Backlight 7 Powering Off 7 Navigation 7 First Power On 8 System Setup Menu 9 DO Local 10 Last...

Страница 3: ...YSI will repair or replace at its sole discretion free of charge any product that YSI determines to be covered by this warranty To exercise this warranty call your local YSI representative or contact...

Страница 4: ...arts or materials are damaged or missing contact YSI Customer Service at 800 897 4151 1 937 767 7241 or the authorized YSI distributor from whom the instrument was purchased BATTERY INSTALLATION The i...

Страница 5: ...he bottom of the Run screen and to increase numerical inputs 3 Power and Backlight Press once to turn instrument on Press a second time to turn backlight on Press a third time to turn backlight off Pr...

Страница 6: ...Next remove the o ring and slide the lid off the sensor INSTALLING THE DISSOLVED OXYGEN SENSOR Before installing a sensor or connecting the cable to the instrument the Sensor Type must be configured f...

Страница 7: ...cap from the dissolved oxygen sensor by pulling it straight off Thoroughly rinse the sensor tip with distilled or deionized water Fill the cap membrane 3 4 full of electrolyte solution then tap the ca...

Страница 8: ...dissolved oxygen sensor type and membrane type Use the up or down arrow keys to highlight the appropriate language sensor and membrane then press enter to confirm figure 6 The Sensor Type must be con...

Страница 9: ...instead of during calibration Example Instrument 1 with DO Local enabled At 737 mmHg barometric pressure roughly 841 ft above sea level the instrument would calibrate to 100 When taking measurements...

Страница 10: ...ompatible sensors for use with a field cable Polarographic This sensor has a black sensor body and is engraved with the model number 2003 Polarographic will be abbreviated Polaro in the instrument men...

Страница 11: ...lay steadily and the instrument will beep twice if Audio is turned on DO Auto Stable can be set to a change of 0 0 to 1 9 over 3 to 19 seconds ISE Auto Stable can be set to a change of 0 0 to 9 9 over...

Страница 12: ...nge the language Highlight the desired language English Spanish Portuguese or French and press enter to enable The enabled language will have an X in the box next to it Highlight the ESC Exit box and...

Страница 13: ...ate formula for this uncorrection is True BP Corrected BP 2 5 Local Altitude in ft above sea level 100 Although the Pro1020 barometer range is 400 0 to 999 9 mmHg you can not adjust the value across t...

Страница 14: ...ubleshooting guide for possible solutions CALIBRATING IN PERCENT DO LOCAL ENABLED Perform this calibration procedure when DO Local is enabled and Quick DO Cal is disabled in the System Setup menu 1 Pe...

Страница 15: ...5 If necessary use the up and down arrow keys to adjust the pH buffer value Note the pH mV reading which ideally should be between 50 and 50 in buffer 7 6 Press enter to continue to second point 7 Ri...

Страница 16: ...membrane If using the 1001A amplified pH sensor be sure to use the guard extender supplied in the 605216 amplified pH sensor kit The extender is screwed onto the sensor bulkhead before the guard or c...

Страница 17: ...enter to regain access to scrolling through the saved data The data set displayed is indicated by the data set number 01 through 50 ERASING DATA While viewing saved data press the enter key to access...

Страница 18: ...or connector Check the o ring and the groove for any excess grease or contamination If contamination is evident clean the o ring and nearby plastic parts with lens cleaning tissue or equivalent lint f...

Страница 19: ...the sensor in clean water wipe with a cotton swab saturated with clean water not DI water and then rerinse with clean water To be certain that all traces of the acid are removed i from the sensor cre...

Страница 20: ...d in the membrane kit Do not sand the electrodes every membrane change as this is not routine maintenance In fact visually the anode may appear tarnished and operate properly YSI recommends using the...

Страница 21: ...embrane should be changed if a bubbles are visible under the membrane b dried electrolyte is visible around the membrane and c if the sensor shows unstable readings or other sensor related symptoms Th...

Страница 22: ...em for service Symptom Possible Solution Instrument will not calibrate pH or ORP instrument displays Calibration Over Calibration Under or Unstable Reading during calibration 1 Verify correct ISE sens...

Страница 23: ...specifications represent typical performance and are subject to change without notice For the latest product specification information please visit YSI s website at ysi com or contact YSI Tech Suppor...

Страница 24: ...E blue cap 6 each includes electrolyte Used with Galvanic sensor All cables include a temperature sensor The pH or ORP and Dissolved Oxygen sensors are sold separately D E C L A R AT I O N O F C O N F...

Страница 25: ...country and region and users are expected to understand and follow the battery disposal requirements for their specific locale C O N TA C T I N F O R M AT I O N ORDERING AND TECHNICAL SUPPORT Telephon...

Страница 26: ...Item 605187 Rev A January 2014 2014 YSI Incorporated...

Отзывы: