Protection
Barrier*
black tip
red tip
insulated conductor
safety cap
metal contact
Ø 2 mm
The internal contact part of the plug is protected by
a safety cap. Remove the cap before use.
plug
*Hold the test leads
with your fingers
protected by the barrier
delimiting the accessible
area.
4.1.1. Procedure for insertion of the batteries
4.2. Test leads
Disconnect the test leads from the dut and from the input terminals of the instrument. Turn the rotary
switch to
OFF
.
Warnings on the test leads
1. Unscrew the two fixing screws of the rear
case.
2. Remove the rear case.
3. Remove the used batteries and replace
them with new ones, after checking the
correct polarity.
4. Reassemble the rear case.
5. Fasten the two fixing screws again.
ATTENTION!
Read the manual!
EN 61010-031 standard prescribes
a maximum length of 4mm for the
metal contact part of the test leads, to prevent
short-circuits or electric arcs during their applica-
tion on the live circuit (CAT III and CAT IV). For this
reason, the test leads supplied with the instrument
are complete with additional removable protec
-
tions (red and black) positioned on the metal
contact part.