Maintenance
Pges
3 - 28
ba43121e03
01/2011
Note
After this service, the TresCon Pges begins with AutoCal as soon as a
new measurement starts.
14
Click the cassette into place in the peristaltic pump and close
the gripping lever.
15
Connect the sample tubing to the overflow vessel again.
16
After performing these tasks, confirm each of the prompts of
2nd part of the checklist with
3
. After the last confirmation,
the 2nd part of the procedure begins immediately. The
remaining time appears on the display.
17
At the end of the procedure,
Service finished
appears on the
display.
18
Return to the service menu with
3
.
Pump cassette
Pump direction
Section 1
(used)
Section 2
Gripping lever
Tighten
Loosen